Table 10.
Results from 28 independent end‐to‐end hidden0target tests, 12 performed using 2D/2D (ExacTrac) localization and 16 performed using cone‐beam CT localization
Hidden Target Offset (mm) | |||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Anterior—Posterior | Right‐Left | Superior—Inferior | |||||||||||||||
Ave | Min | Max | STD | Ave | Min | Max | STD | Ave | Min | Max | STD | ||||||
All Data | 0.00 |
|
0.79 | 0.34 |
|
|
0.46 | 0.27 | 0.19 |
|
0.73 | 0.27 | |||||
2D/2D | 0.14 |
|
0.79 | 0.37 |
|
|
0.46 | 0.29 | 0.24 |
|
0.69 | 0.27 | |||||
CBCT |
|
|
0.48 | 0.28 |
|
|
0.06 | 0.27 | 0.15 |
|
0.73 | 0.27 |