Skip to main content
. 2014 Mar 6;15(2):181–190. doi: 10.1120/jacmp.v15i2.4647

Table 2.

Mean three‐dimensional gamma agreement index (GAI) and standard deviation inside the PTV and the body of the patient using Delat4 Anatomy option. Mean GAI in the horizontal isocenter slice and mean dose deviation in the isocenter and mean GAI in the phantom (Conventional GAI)

Prostate Brain
Body GAI (3mm/5%)
94.0%±2.9%
97.9%±1.5%
PTV GAI (3mm/5%)
97.3%±2.0%
94.4%±1.7%
Body GAI (3mm/3%)
86.0%±5.0%
95.3%±2.1%
PTV GAI (3mm/3%)
83.7%±8.8%
90.6%±2.3%
Horizontal slice (3mm/3%)
87.2%±5.1%
95.2%±2.3%
Dose deviation (%)
2.4%±2.5%
1.3%±2.1%
Conventional GAI (3mm/3%)
99.8%±0.4%
99.7%±0.4%