Skip to main content
. 2014 Mar 6;15(2):110–120. doi: 10.1120/jacmp.v15i2.4620

Table 1.

Parameters and results for hand calculations of DVH statistics using the 125I Oncura Model 6711 source. Constants used in the calculation include: SK(U)=1.01,Λ(cGy/Uh)=0.965,ϕan(r)=1, and τ(h)=2053

r (cm)
(r0/r)2
gp(r)
D(Gy)
%VolTG43
%DoseTG43
1 1.000 1.000 20.01 19.1 100.0
2 0.250 0.819 4.10 39.2 20.5
3 0.111 0.636 1.41 59.2 7.1
4 0.063 0.499 0.62 79.2 3.1
5 0.040 0.367 0.29 99.2 1.5