Table 4.
Mean dose values for planning target volumes (PTV) and organs at risk (OAR) for the evaluated left chest wall plans, and for each of the detector‐specific models. The percent difference from the CC13 values are calculated by subtracting the tested values by the CC13 values, and normalized to 100% using the CC13 values
CC13 | 60012 | % diff from CC13 | EDGE | % diff from CC13 | |||
---|---|---|---|---|---|---|---|
Left Chest Wall Plan 1 | |||||||
Mean Dose (cGy) | PTV | 4543 | 4538 |
|
4543 | 0.0 | |
Left Lung | 203 | 207 | 2.0 | 203 | 0.0 | ||
Heart | 115 | 121 | 5.4 | 115 |
|
||
Left Chest Wall Plan 2 | |||||||
Mean Dose (cGy) | PTV | 5102 | 5096 |
|
5100 | 0.0 | |
Left Lung | 560 | 566 | 1.0 | 560 |
|
||
Heart | 107 | 114 | 6.8 | 107 |
|
||
Left Chest Wall Plan 3 | |||||||
Mean Dose (cGy) | PTV | 4466 | 4459 |
|
4466 | 0.0 | |
Left Lung | 490 | 494 | 0.9 | 490 |
|
||
Heart | 98 | 105 | 7.6 | 99 |
|