Skip to main content
. 2014 Jul 8;15(4):174–193. doi: 10.1120/jacmp.v15i4.4801

Table 6.

Mean dose values for planning target volumes (PTV) and organs at risk (OAR) for the static‐field plans, and for each of the detector‐specific models. The percent difference from the CC13 values are calculated by subtracting the tested values by the CC13 values, and normalized to 100% using the CC13 values

CC13 60012 % diff from CC13 EDGE % diff from CC13
Wedged Pair Plan
Mean Dose (cGy) PTV 1447 1445
0.1
1447 0.0
Brain Stem 630 630 0.2 630 0.0
Cord 734 735 0.2 733 0.0
Lt. Parotid 60 64 0.3 60 0.0
AP/PA Plan 1
Mean Dose (cGy) Larynx (PTV) 6529 6530 0.0 6530 0.0
Cord 67 71 7.2 67
0.2
AP/PA Plan 2
Mean Dose (cGy) Spleen (PTV) 517 516
0.1
517 0.6
Cord 350 350 0.2 350 0.7
Lt. Kidney 503 504 0.2 503 1.2
Rt. Kidney 117 117 0.3 117 0.4