Table 3.

Phantom temperatures (mean ± SD) measured via optic fiber sensors and times to equilibrate with temperature inside a magnet bore at 3 T

Temperature inside magnet bore = 22.8 ± 0.6°C
Sensor No. Initial temperature Final temperature Time to equilibrate with the environment
OFS#1 28.4 ± 0.5°C 22.7 ± 0.6°C 4 hrs 08 min
OFS#2 28.3 ± 0.6°C 22.8 ± 0.5°C 4 hrs 02 min
OFS#3 28.3 ± 0.6°C 22.8 ± 0.6°C 4 hrs 07 min
OFS#4 28.2 ± 0.5°C 22.7 ± 0.6°C 4 hrs 11 min