Skip to main content
. Author manuscript; available in PMC: 2019 Mar 2.
Published in final edited form as: Phys Med Biol. 2018 Mar 2;63(5):055008. doi: 10.1088/1361-6560/aaad76

Table 1.

Summary of mean CRC values for 5 and 10 mm lesions at radial positions of 7 and 13 cm in the lesion phantoms.

Sphere diameter (mm) Radial position (cm) Anger positioning 1st max crystal positioning 2nd max crystal positioning
10 7 0.56 0.50 0.64
10 13 0.53 0.48 0.62
5 7 0.17 0.16 0.22
5 13 0.15 0.13 0.18