Published in final edited form as: Magn Reson Med. 2017 Nov 16;80(1):371–379. doi: 10.1002/mrm.27007

Figure 8.

Figure 8

EPI images corresponding to the B0 maps shown in Figure 4. Perturbation loops were used to introduce B0 inhomogeneities and the iPRES(2)-A coil was used for both imaging and B0 shimming. EPI images with no perturbation (a,e); with the single loop or figure-eight perturbation before shimming (b,f), showing severe geometric distortions (red arrows); after shimming with the optimal switch state and DC current applied to the iPRES(2)-A coil (c,g), showing a drastic reduction in distortions (green arrows); and after shimming with the iPRES(1)-equivalent switch state (d,h), showing only partial or minimal reduction in distortions (orange and red arrows).