Table 2.

Mean deviations (in dB) of hearing aid gain measured in HA2-2cc coupler from prescriptive gain targets at each frequency for low (50 dB SPL), medium (65 dB SPL), and high (80 dB SPL) input levels.

Frequency (Hz)
Input level 250 500 1000 2000 3000 4000
Low (50 dB SPL) Mean −2.0 −1.4 −1.2 0.8 −1.7 −0.9
SD 4.2 3.8 3.5 3.7 4.1 4.6
Medium (65 dB SPL) Mean −0.9 −0.1 0.0 −0.1 −2.0 −1.0
SD 3.6 3.3 3.1 3.3 4.1 4.7
High (80 dB SPL) Mean −1.7 −0.5 0.5 0.9 0.1 1.0
SD 4.3 3.4 3.4 3.6 4.8 5.1