Skip to main content
. 2018 Apr 4;19:104. doi: 10.1186/s12891-018-2026-6

Table 2.

Comparative analysis of screw placement accuracy in 2 groups

group Screw parallelism (deg.) Screw area ratio DS (mm) DC (mm) DSC (mm)
AP Lateral
conventional 1.73 ± 0.62 1.52 ± 0.60 0.16 ± 0.02 10.38 ± 3.91 8.02 ± 1.51 4.21 ± 1.31
GFNS 1.68 ± 0.83# 1.61 ± 0.57# 0.29 ± 0.02* 16.87 ± 4.12* 11.05 ± 1.94* 2.40 ± 1.24*

GFNS guide of femoral neck section, DS distance between screws, DC distance between the centre of FNS, DSC distance between screw to femoral neck cortex

*P < 0.05; #p > 0.05 Comparing with Conventional group