Table 3.

B0 eddy field measured for the unshielded gradient coil. The amplitude is normalized by the amplitude for the gradient field switching of 1 T/m/s

Gx Gy Gz



Component 1 Component 2 Component 1 Component 2 Component 1 Component 2
Time constant [ms] 0.84 11.6 0.084 12.6 0.048 7.26
Amplitude [μT/(T/m)] 183 −7.4 −1766 7.68 128 65.8