Skip to main content
. 2016 Sep 19;1:2–8. doi: 10.1016/j.ctro.2016.08.001

Table 5.

Treatment delivery metrics.

Target Unit Autoplan
Manual
p
Mean STD Mean STD
MU MU 435 79 360 56 <0.001
Operator time min 32 26 64 31 <0.001
Beam on time s 114 22 110 24 0.017
Pass rate (3%,3 mm) % 97.7 1.7 98.4 1.7 0.001
Pass rate (2%,2 mm) % 90.6 4.3 92.5 4.5 0.003