Table 3. Comparison data of four carcinoembryonic antigen (CEA) immunoassays using 355 serum samples with CEA<100 ng/mL.

Statistics Centaur XP (y) vs i2000sr (x) Centaur XP (y) vs E170 (x) Centaur XP (y) vs DxI800 (x) i2000sr (y) vs E170 (x) i2000sr (y) vs DxI800 (x) E170 (y) vs DxI800
Passing-Bablok regression
 Regression equation y=−0.629+0.628x y=−0.377+0.671x y=−0.038+0.709x y=0.665+1.038x y=1.214+1.114x y=0.557+1.045x
 Intercept (95% CI) −0.910 to −0.386 −0.595 to −0.181 −0.199 to 0.104 0.269 to 1.048 0.640 to 1.727 0.240 to 0.904
 Slope B (95% CI) 0.597 to 0.657 0.648 to 0.693 0.691 to 0.729 0.996 to 1.074 1.049 to 1.175 1.002 to 1.085
 Cusum test for linearity ns (P=0.97) ns (P=0.88) ns (P=0.20) ns (P=0.31) ns (P=0.46) ns (P=0.38)
Bland-Altman analysis
 Mean difference (ng/mL) (95% CI) −8.1 (−9.1 to −7.2) −6.5 (−7.2 to −5.7) −5.1 (−6.0 to −4.3) 1.6 (1.1 to 2.2) 3.0 (2.2 to 3.8) 1.3 (0.6 to 2.1)
P value <0.0001 <0.0001 <0.0001 <0.0001 <0.0001 0.0003

Abbreviations: y, y axis; x, x axis; 95% CI, 95% confidence interval.