Skip to main content
. Author manuscript; available in PMC: 2018 Apr 25.
Published in final edited form as: Circ J. 2017 Feb 2;81(5):682–688. doi: 10.1253/circj.CJ-16-1109

Table 5.

Exclusion Criteria for IIA preserving devices in 87 aorto-iliac segments

Cook IBD Exclusion Criteria n (%) Gore IBE Exclusion Criteria n (%)
CIA length < 50 mm 34 (39) Aorto-iliac length < 165 mm 21 (24)
CIA diameter < 20 mm 2 (2) CIA diameter < 25 mm 5 (6)
EIA length < 20 mm 8 (9) Proximal CIA diameter < 17 mm 39 (44)
EIA diameter < 8 mm 32 (37) Distal CIA diameter < 14 mm 4 (5)
IIA occluded or 50% stenosis 1 (1) EIA Length < 10 mm 6 (7)
IIA aneurysm distal to landing zone 9 (10) EIA diameter <6.5 or > 25 mm 13 (15)
IIA length < 10 mm 0 (0) IIA length < 10 mm 0(0)
IIA diameter < 6 or > 9 mm 41 (47) IIA diameter < 6.5 or > 13.5 mm 13 (15)
    Any Factor 72 (83)     Any Factor 65 (75)

IBD - Iliac branch device, IBE - Iliac branch endoprosthesis