Ablation protocol for 5-cm ablation (RITA Medical, Boston Scientific).

Deploy to Target temp Power Timer Heating duration
2 cm 105 °C 150 W 15 min Until target temp. is reached, then deploy to 3 cm
3 cm 105 °C 150 W 14.5 min Until target temp. is reached, then deploy to 4 cm
4 cm 105 °C 150 W 14 min 7 min after target temp (i.e., after hearing the beep, wait 7 min), then deploy to 5 cm