The DESS-MS sequence served as the morphometry reference sequences while the DESS-T2S served as the relaxometry reference sequence. A multi-echo fast-spin-echo sequence with eight echoes was also used for T2 comparisons (*meniscus T2 measurements utilized TEs of 6ms, 13ms, 19ms, and 26ms while cartilage T2 mapping utilized additional TEs of 32ms, 38ms, 45ms, and 51ms). All DESS sequences had a spoiler gradient area of 31.3 mT1m−1ms1 (1.33 cycles of dephasing/mm) and the minimum achievable TE and TR for the specified resolution and bandwidth. Note that for the clinical implementation of DESS-EXP, a readout of 416 samples and a slice thickness of 1.6mm was used to automatically reconstructed 512x512 images and to interpolate 0.8mm thick slices to 2.4mm thicknesses. Abbreviations: RO = readout, PE = phase encodes.