Published in final edited form as: J Magn Reson Imaging. 2017 Nov 1;47(5):1328–1341. doi: 10.1002/jmri.25883

Table 1.

The DESS-MS sequence served as the morphometry reference sequences while the DESS-T2S served as the relaxometry reference sequence. A multi-echo fast-spin-echo sequence with eight echoes was also used for T2 comparisons (*meniscus T2 measurements utilized TEs of 6ms, 13ms, 19ms, and 26ms while cartilage T2 mapping utilized additional TEs of 32ms, 38ms, 45ms, and 51ms). All DESS sequences had a spoiler gradient area of 31.3 mT1m−1ms1 (1.33 cycles of dephasing/mm) and the minimum achievable TE and TR for the specified resolution and bandwidth. Note that for the clinical implementation of DESS-EXP, a readout of 416 samples and a slice thickness of 1.6mm was used to automatically reconstructed 512x512 images and to interpolate 0.8mm thick slices to 2.4mm thicknesses. Abbreviations: RO = readout, PE = phase encodes.

# 1 2 3 4
Sequence DESS DESS ME-FSE DESS
Purpose Morphometry Standard T2 Standard T2 Comparison Test Experimental
Shorthand Name DESS - MS DESS - T2S ME-FSE DESS - EXP
Matrix (RO × PE) 512×512 256×256 128×128 384×512
Resolution (mm2) 0.31×0.31 0.62×0.62 1.25×1.25 0.42×0.31
Slice Thickness (mm) 1.5 3 3 1.5
Number of Slices 80 40 20 80
TE S+ (ms) 6.5 5.1 N/A* 5.7
TE S(ms) 35.5 25.5 N/A* 30.1
Number of Echoes 2 2 8 2
TR (ms) 21 15.3 1650 17.9
Flip Angle (°) 20 20 90 20
Bandwidth (± kHz) 42 31 31 42
Parallel Imaging None None None 2x1
% Corners Cut 0 0 0 25
Scan Time (mm:ss) 14:20 2:37 2:30 4:48