Skip to main content
. 2018 Apr 1;2018:8274131. doi: 10.1155/2018/8274131

Table 2.

Comparison of the GC-MS [8] versus LC-MS/MS method.

Criterion GC-MS LC-MS/MS
Retention time PHT 15.12 min, IS MPPH 16.15 min PHT and PHT-D10 2.8 min
Selectivity/sensitivity (absence of PHT) Good peak differentiation and quantification of PHT. All blank samples were negative (no presence of PHT) All blank samples were negative (no presence of PHT)
Recovery 94.1% for QC2 (100 ng/mL) 89.5% for QC1 (10 ng/mL)
94.3% for QC5 (1000 ng/mL) 97.1% for QC3 (1600 ng/mL)
LOD (calculated as S/N ratio of 4 : 1) 15 ng/mL <1 ng/mL
LOQ (calculated as 5 times the response/blank) 50 ng/mL 10 ng/mL
Accuracy The calibrator values showed min–max percent deviations of 1–20% for Cal 1 (50 ng/mL) to Cal 6 (1200 ng/mL) The calibrator values showed min–max percent deviations of 1–8% for Cal 1 (10 ng/mL) to Cal 8 (2000 ng/mL)
Regression coefficient, r2 r 2 blood = 0.998 (n = 2)
r2dialysate = 0.999 (n = 8)
r2saliva = 0.999 (n = 2)
r 2 blood = 0. 996 (n = 3)
r2dialysate = 0.997 (n = 6)
r2saliva = 0.995 (n = 3)
Calibration range 50–1200 ng/mL 10–2000 ng/mL
Run time per analysis 30 min 7 min
Injection volume of the sample 2.0 µL 10 µL
Sample preparation time 5 h for 25 samples 6 h for 182 samples
Stability of the processed samples Dried extracts were stable for ≥4 weeks (min/max deviation 4%). No effect by reinjection and storage (33 h) on the autosampler Reinjection after 7 days showed no difference in accuracy
Sample volume needed 50 µL 25 µL