Skip to main content
. 2018 Mar 15;45(4):1444–1458. doi: 10.1002/mp.12812

Table 3.

Head phantom DEQC protocol version 1

Acquisitions GSI‐preset Pitch Rotation time (s) mA CTDIvol (mGy) Image thickness (mm)
1, 2 GSI‐26 0.531 0.7 375 67 1.25, 3.75
3, 4 GSI‐26 0.969 0.7 375 36.7 1.25, 3.75
5, 6 GSI‐20 0.531 0.5 630 81.4 1.25, 3.75
7, 8 GSI‐20 0.969 0.5 630 44.6 1.25, 3.75
9, 10 GSI‐9 0.531 0.9 600 132.6 1.25, 3.75
11, 12 GSI‐9 0.969 0.9 600 72.7 1.25, 3.75

Note: All acquisitions used helical scan mode, beam width of 20 mm, Head SFOV, DFOV of 25 cm, and Standard reconstruction algorithm. A total of five images were acquired per acquisition.