Table 8.
Test metric | Image type | Units | CTDIvol | Rotation time | Pitch | Image thickness |
---|---|---|---|---|---|---|
Iodine quant error 2 mg/mL | I(Wa) | mg/mL, % | 0.22, 10.77% | 0.11, 5.59% | 0.03, 1.33% | 0.02, 1.23% |
Iodine quant error 5 mg/mL | I(Wa) | mg/mL, % | 0.32, 6.35% | 0.18, 3.70% | 0.04, 0.85% | 0.00, 0.01% |
Iodine quant error 15 mg/mL | I(Wa) | mg/mL, % | 0.16, 1.06% | 0.10, 0.65% | 0.02, 0.12% | 0.16, 0.05% |
Monoenergetic HU stability | 50 keV | HU | 14.45 | 8.26 | 0.16 | 0.11 |
70 keV | HU | 2.1 | 1.3 | 0.22 | 0.07 | |
110 keV | HU | 5.79 | 2.86 | 0.4 | 0.04 | |
140 keV | HU | 7.43 | 3.78 | 0.44 | 0.04 | |
Noise | 50 keV | HU | 34.35 | 10.78 | 11.48 | 10.36 |
70 keV | HU | 16.51 | 5.06 | 6.22 | 5.62 | |
110 keV | HU | 14.47 | 4.56 | 4.77 | 4.22 | |
140 keV | HU | 13.55 | 4.35 | 4.32 | 3.76 | |
I(Wa) | mg/mL | 5.43 | 1.77 | 1.61 | 1.39 | |
Wa(I) | mg/mL | 12.81 | 4.22 | 3.81 | 3.21 | |
Positional constancy | 50 keV | HU | 10.26 | 9.24 | 0.57 | 0.52 |
70 keV | HU | 0.89 | 0.22 | 0.25 | 0.12 | |
110 keV | HU | 5.76 | 5.67 | 0.03 | 0.44 | |
140 keV | HU | 7.04 | 6.91 | 0.08 | 0.52 | |
I(Wa) | mg/mL | 0.36 | 0.34 | 0.01 | 0.02 | |
Wa(I) | mg/mL | 8.85 | 8.68 | 0.14 | 0.67 |
Note: Values in bold were identified as statistically significant while those underlined were identified as both statistically and nominally significant.