Table 8.

Maximum impact of technique parameter on test metrics for the DEQC body phantom (Protocol 1)

Test metric Image type Units CTDIvol Rotation time Pitch Image thickness
Iodine quant error 2 mg/mL I(Wa) mg/mL, % 0.22, 10.77% 0.11, 5.59% 0.03, 1.33% 0.02, 1.23%
Iodine quant error 5 mg/mL I(Wa) mg/mL, % 0.32, 6.35% 0.18, 3.70% 0.04, 0.85% 0.00, 0.01%
Iodine quant error 15 mg/mL I(Wa) mg/mL, % 0.16, 1.06% 0.10, 0.65% 0.02, 0.12% 0.16, 0.05%
Monoenergetic HU stability 50 keV HU 14.45 8.26 0.16 0.11
70 keV HU 2.1 1.3 0.22 0.07
110 keV HU 5.79 2.86 0.4 0.04
140 keV HU 7.43 3.78 0.44 0.04
Noise 50 keV HU 34.35 10.78 11.48 10.36
70 keV HU 16.51 5.06 6.22 5.62
110 keV HU 14.47 4.56 4.77 4.22
140 keV HU 13.55 4.35 4.32 3.76
I(Wa) mg/mL 5.43 1.77 1.61 1.39
Wa(I) mg/mL 12.81 4.22 3.81 3.21
Positional constancy 50 keV HU 10.26 9.24 0.57 0.52
70 keV HU 0.89 0.22 0.25 0.12
110 keV HU 5.76 5.67 0.03 0.44
140 keV HU 7.04 6.91 0.08 0.52
I(Wa) mg/mL 0.36 0.34 0.01 0.02
Wa(I) mg/mL 8.85 8.68 0.14 0.67

Note: Values in bold were identified as statistically significant while those underlined were identified as both statistically and nominally significant.