Table 10.

Recommendations for long‐term data collection to determine baseline values and tolerances necessary for the establishment of a DECT quality control program. All acquisitions apply to the DEQC body phantom and are performed in helical scan mode, using a 40 mm beam width, Large Body SFOV, 42 cm DFOV, and standard reconstruction filter

Acquisition GSI‐preset Pitch Rotation time (s) mA Image thickness (mm) CTDIvol (mGy)
1 GSI‐1 0.516 0.5 630 5 33.9
2 GSI‐5 0.984 1 600 5 32.5
3 GSI‐10 0.984 0.8 600 5 25.5
4 GSI‐1 0.984 0.5 630 5 17.8
5 GSI‐54 0.516 0.6 275 5 17.4
6 GSI‐48 0.516 0.7 260 5 17.2