. 2018 Mar;10(3):2059–2069. doi: 10.21037/jtd.2018.03.103

Table 2. Efficiency of the test mask without Micro Ventilator for filtration of Staphylococcus aureus and bacteriophage ΦΧ174.

Replicate % Filtration assessed by infectivity
Staphylococcus aureus Bacteriophage ΦΧ174
Replicate 1 >99.9% 99.6%
Replicate 2 >99.9% 99.8%
Replicate 3 99.7% >99.9%
Replicate 4 99.8% 99.8%
Replicate 5 >99.9% 99.5%
Mean ± SD (N=5 replicates) 99.8%±0.1%* 99.7%±0.2%*

This work was performed by Nelson Laboratories (Salt Lake City, USA). *, For the purpose of averaging and statistical analysis, values indicated as >99.9% were treated as 99.9%. SD, standard deviation.