Table 3. Efficiency of the test mask with Micro Ventilator for filtration of particulates.

Conditioning of mask per EN149 % Penetration (mean ± SD, N=3)
Sodium chloride (solid) Paraffin oil (liquid)
As received 99.9%±0.0% 99.7%±0.0%
Simulated wearing 99.9%±0.0%* 99.4%±0.1%*
Mechanical strength and temperature conditioning 99.9%±0.0% 99.6%±0.1%
Overall (N=9) 99.9%±0.0% 99.6%±0.1%

This work was performed by INSPEC (Greater Manchester, UK). *, significantly different from the “as received” value by two-tailed t-test. SD, standard deviation.