Overall percentage reduction in periprosthetic stress and strain for femurs implanted with stemmed implants relative to a PS implanted femur

Model Stem length (mm) Decrease in von Mises stress (%) Decrease in equivalent strain (%)
PS N/A 0.0 0.0
TS short 50 10.9 10.0
TS medium 75 25.7 25.5
TS long 100 28.7 29.3

Where for each femur, values of stress and strain are calculated based on the average of the four cortical points of interest