Figure 5.

Figure 5

A) Image of lung tissue in which Focal CC was repeatedly cycled off and on during a single continuous frame acquisition (∼90 seconds) showing rapid and complete charge elimination and no residual effect of gas injection on focus or stigmation. The black horizontal bands represent massive charging that occurs with the gas cycled off. B and C are enlargements of the boxed regions in A. Bars = 500 nm.