Published in final edited form as: Clin Neurophysiol. 2017 Oct 21;129(1):308–318. doi: 10.1016/j.clinph.2017.10.004

Table 3.

True and false ripple on spike detector sensitivity, specificity, precision i.e. positive predictive value (PPV), and negative predictive value (NPV) for differentiating the two event types in the 12 study patients in a blinded review by two experts conducted without spectrogram display.

Patient ID Accuracy Sensitivity Specificity Precision (PPV) nPV Kappa K. err
448 (n = 50) 0.78/0.72 0.72/0.75 0.84/0.64 0.82/0.69 0.75/0.76 −0.14 0.14
449 (n = 62) 0.71/0.58 0.97/0.52 0.45/0.65 0.64/0.59 0.94/0.57 0.03 0.12
451 (n = 272) 0.82/0.66 0.89/0.94 0.74/0.38 0.78/0.6 0.87/0.87 0.07 0.07
453 (n = 86) 0.71/0.57 0.84/0.74 0.58/0.4 0.67/0.55 0.78/0.61 −0.12 0.12
454 (n = 210) 0.74/0.74 0.71/0.85 0.76/0.63 0.75/0.70 0.73/0.81 0.08 0.07
456 (n = 192) 0.71/0.65 0.82/0.79 0.60/0.50 0.68/0.61 0.77/0.71 −0.10 0.08
463 (n = 116) 0.57/0.62 0.60/0.66 0.53/0.59 0.57/0.61 0.57/0.63 −0.12 0.09
467 (n = 108) 0.80/0.71 0.78/0.70 0.82/0.72 0.81/0.72 0.79/0.71 0.07 0.10
470 (n = 72) 0.61/0.63 0.67/0.81 0.56/0.44 0.60/0.59 0.63/0.70 −0.04 0.12
472 (n = 36) 0.56/0.50 0.50/0.44 0.61/0.56 0.56/0.50 0.55/0.50 −0.08 0.17
473 (n = 72) 0.71/0.67 0.58/0.75 0.83/0.58 0.78/0.64 0.67/0.70 0.25 0.11
474 (n = 94) 0.82/0.72 0.89/0.89 0.89/0.55 0.78/0.67 0.88/0.84 0.02 0.11
All Patients (n = 1370) 0.73/0.66 0.75/0.74 0.67/0.55 0.70/0.62 0.74/0.70 0.01 0.03