TABLE 2.

Troubleshooting table

Step Problem Possible reason Solution
9 Too many bubbles around worms Too little anesthetic liquid left on slide Stop removing anesthetic with the mouth pipette once the liquid no longer easily fills the capillary glass without applying suction
Worms too widely distributed on the agarose pad Too much anesthetic liquid is left on the pad Try to minimize the amount of liquid transferred with the worms by gently tapping the glass spot plate (the worms will coalesce at the bottom of the well (see Fig. 3c), allowing you to draw them up in a minimal amount of liquid)
10 Worms become sick or die during imaging Too much OP50 is transferred with worms
Worms are injured while they are anesthetized, transferred, and covered by a coverslip
The interface between the coverslip and the agarose pad is not properly sealed
Use a minimal amount of bacteria to transfer worms
Use care when handling worms at all points
Use a small paintbrush to apply VALAP to four sides of the coverslip and leave a small gap for gas exchange
14 Worms are not at the correct stage in development to image Wrong sized animals are picked
Worms cultured for too short or long after syncing
Worms not kept at an optimal temperature
Sync animals by bleaching instead of trying to pick from mixed aged plates
Optimize the correct incubation time and temperature for each strain
15 Specimens with low fluorescence signal will result in grainy images EM Gain might be too high while exposure time is likely too short Adjust the settings until high-quality images are acquired
Worms start moving during time-lapse Worms are too crowded or the incubation in anesthetic was too short Lengthen incubation time in anesthetic, avoid mounted worms that are touching each other, as they tend to move more during time-lapses
Photobleaching occurs during imaging Worms are exposed to too much cumulative laser power Reduce laser power, exposure time
17 Imaging data cannot be loaded into Imaris The format of data files is not properly selected in Imaris Try to correctly match file format selected in Imaris with that of raw data files
18 Disproportionate 3D rendering Geometry set in Imaris does not match acquired raw images Verify correct voxel dimensions. If voxel size was determined by calculation, confirm with a stage micrometer