Table 1. Within-run precision.

The mean cell number and coefficient of variations (% CV) of ten different CSF samples were calculated five consecutive times with the XN-hsA and the XN-BF modes.

XN-hsA mode XN-BF mode
Sample# Mean (cells/μl) CV (%) Sample# Mean (cells/μl) CV (%)
1 256.7 5.6 6 333 2.6
2 19.0 5.7 7 100.2 5.3
3 6.2 7.5 8 28.6 8.0
4 1.3 13.6 9 6.8 12.3
5 0.5 24.8 10 1.2 37.3