Skip to main content
. Author manuscript; available in PMC: 2019 Apr 20.
Published in final edited form as: Phys Med Biol. 2018 Apr 20;63(8):085019. doi: 10.1088/1361-6560/aab83b

Table 1.

Parameters considered in photon budget and typical range of quanta available at each stage.

Component Parameters Parameter range Quanta range (N/mm2)
Radiation γ (MV) or e (MeV),
Dose (Gy),
Rate (Gy/s)
6–18(MV), 6–18(MeV),
0–2(Gy),
0.017 – 0.1 (Gy/s)
Cherenkov β,
n
0.81–0.98,
1.3–1.5
1.0E+07–2.0E+11
Tissue μ-environment
expd0μa(mm1)
d0: <1mm
μa: 0.02–20.0 (mm−1)
8.0E+06–2.0E+11
Luminescence Φ (%),
τ (μs)
0.01–0.3,
1–300 (μs)
8.0E+04–6.0E+10
Tissue diffusion
expd1μeff(mm1)
d1: 1–20 (mm)
μeff: 0.1–10.0 (mm−1)
5.0E+02–5.0E+07
Camera placement
1d22
d2: 300–2000 (mm) 1.0E−04–6.0E+02
Lens
2sin1(NA)π
NA: 0.09–0.42 7.0E−06–2.0E+02
Signal Background Signal Background Signal Background
Photocathode QE(λ)
(%)
EBI
(ph/pix/s)
0.01–0.35 0.002–0.02 7.0E−08–6.0E+01 8.0E−08–1.0E−04
MCP Gain 104–107 7.0E−04–6.0E+08 8.0E−04–1.0E+03
Phosphor plate η
(ph/e)
τ (ms) 95–200 0.2μs–3ms 7.0E−02–1.0E+11
Coupling 50% 30–80 % 3.0E−02–6.0E+10
CCD QE(λ)
(%)
σS, σD 0.3–0.5 N,
<2
1.0E−02–2.0E+10
ADC Rate (MHz) Read noise (e) 2–32 10–70 (e)