. 2018 Apr 27;18:484. doi: 10.1186/s12885-018-4400-z

Table 2.

Results obtained from samples of the round robin test. Reference values were obtained from the trial coordinators by a multiplex assay analyzed on a capillary sequencer and reported to our laboratory. Afterwards, the samples were re-analyzed on the QiaXcel Advanced system. All samples could be fully confirmed by analyzing the peak patterns

Round Robin Sample BAT25 BAT26 D5s346 D2S123 D17s250 MS-status
1 expected wt wt wt wt wt MSS/MSI-L
our result wt wt wt wt wt MSS
2 expected MSI MSI MSI MSI MSI MSI-H
our result MSI MSI MSI MSI MSI MSI-H
3 expected wt wt wt wt wt MSS/MSI-L
our result wt wt wt wt wt MSS
4 expected MSI MSI MSI MSI MSI MSI-H
our result MSI MSI MSI MSI MSI MSI-H
5 expected wt wt wt wt wt MSS/MSI-L
our result wt wt wt wt wt MSS
6 expected wt wt wt wt wt MSS/MSI-L
our result wt wt wt wt wt MSS
7 expected MSI MSI MSI MSI MSI MSI-H
our result MSI MSI MSI MSI MSI MSI-H
8 expected MSI MSI MSI MSI MSI MSI-H
our result MSI MSI MSI MSI MSI MSI-H