. 2018 Apr 24;12:747–753. doi: 10.2147/OPTH.S164175

Table 2.

Postoperative visual outcomes and toric IOL alignment in cases undergoing toric IOL implantation with manual marking or image-guided surgery

Outcome measures Group I
(manual marking)
N=40
Group II
(Callisto assisted)
N=40
p-value
Refractive sphere 0.39±0.44 0.15±0.79 0.105
Refractive cylinder −0.89±0.35 −0.64±0.36 0.003
Corneal cylinder 2.1±0.9 2.3±1.2 0.474
Defocus equivalent 0.01±0.2 0.07±0.3 0.208
UDVA (logMAR units) POD 1 0.055±0.09 0.03±0.07 0.184
CDVA (logMAR units) POD 1 0.01±0.03 0.005±0.02 0.402
UDVA (logMAR units) POD 30 0.025±0.06 0.017±0.05 0.541
CDVA (logMAR units) POD 30 0.008±0.03 0.005±0.02 0.649
IOP (mmHg) 15.2±2.9 14.4±2.9 0.272
Axis of IOL implantation 82.6±72.3 104.6±69.0 0.169
Deviation from target axis (POD 1) 5.8±3.7 3.7±2.8 0.005
Deviation from target axis (POD 30) 5.5±3.3 3.6±2.6 0.005

Note: p-values less than 0.05 are shown in bold.

Abbreviations: D, Diopter; UDVA, uncorrected distance visual acuity; CDVA, corrected distance visual acuity; logMAR, logarithm of minimum angle of resolution; IOP, intraocular pressure; IOL, intraocular lens; POD, postoperative day.