Skip to main content
. 2018 Apr 25;56(5):e00050-18. doi: 10.1128/JCM.00050-18

TABLE 2.

Sensitivity of the ICT for the detection of different carbapenemases directly from Mueller-Hinton agar

Inoculum (μl)a Carbapenemase detectedb
OXA-48 like
KPC
NDM
Sensitivity (% [no. detected/total no.]) 95% CI (%) Sensitivity (% [no. detected/total no.]) 95% CI (%) Sensitivity (% [no. detected/total no.]) 95% CI (%)
1 (no AB) 100 (25/25) 86.3–100 100 (10/10) 69.2–100 94.4 (17/18) 72.7–99.9
10 (no AB) 100 (25/25) 86.3–100 100 (10/10) 69.2–100 100 (18/18) 81.5–100
1 EPM 100 (25/25) 86.3–100 100 (10/10) 69.2–100 100 (18/18) 81.5–100
1 IPM 100 (25/25) 86.3–100 100 (10/10) 69.2–100 100 (18/18) 81.5–100
1 MER 100 (25/25) 86.3–100 100 (10/10) 69.2–100 100 (18/18) 81.5–100
a

no AB, inoculum taken distant from antibiotic disk, with all others harvested next to an antibiotic disk; EPM, ertapenem; IPM, imipenem; MER, meropenem.

b

95% CI, 95% confidence interval.