TABLE 6.

Neutralization results for Elecsys Chagas assay-discrepant, repeatedly reactive samples from blood donors

Study site Comparator COI Elecsys COIa Elecsys COI after neutralization Recoveryb (%)
Italy 0.016 2.94 1.87 64
Colombia 0.023 1.16 0.298 26
Colombia 0.037 1.09 0.300 28
Colombia 0.554 1.47 0.515 35
Argentina 0.230 12.6 0.348 3
Argentina 0.99/0.97/1.14c 40.8 2.04 5
Argentina 0.040 2.92 0.227 8
Argentina 0.080 1.60 0.132 8
Argentina 0.510 1.47 0.152 10
Argentina 0.030 1.03 1.01 98
a

COI as determined by Roche Diagnostics Centralized and Point of Care Solutions (Penzberg, Germany) prior to the neutralization procedure.

b

A recovery of ≤25% was assessed as successful neutralization (i.e., positive for anti-T. cruzi antibodies). Values in boldface type represent neutralizable/borderline and neutralizable samples.

c

The COI was repeatedly in the gray zone.