Table 3. Precision and accuracy of the determination of cannabinoids in plant material (Intra-day n = 5).

Analyte Nominal Concentration (ng/mL) Measured Concentration (ng/mL) (mean +/- SEM) Accuracy (%) RSD (%)
THC
Intra-day 0.5 0.50 +/- 0.03 100.4 13.20
5 5.23 +/- 0.07 104.5 3.42
50 52.82 +/- 0.67 105.6 3.08
200* 209.67 +/- 3.83 104.8 4.47
CBD
Intra-day 0.5 0.52 +/- 0.02 103.9 11.51
5 5.11 +/- 0.10 102.3 4.61
50 50.77 +/- 0.20 101.5 0.95
200* 216.24 +/- 3.01 108.1 3.41
CBDA
Intra-day 0.5 0.91 +/- 0.09 182.3 25.43
5 6.87 +/- 0.15 137.4 5.38
50 78.98 +/- 4.35 158.0 13.50
200* 229.47 +/- 6.90 114.7 7.38
THCA-A
Intra-day 0.5 0.73 +/- 0.02 145.3 6.00
5 7.25 +/- 0.35 145.0 11.98
50 103.17 +/- 3.79 206.3 9.00
200* 287.33 +/- 8.50 143.7 7.24

* Quality control samples of 200 ng/mL were diluted 1/10 with extraction solvent before injection.