Table 1.
Reflectivity Changes during Fabrication of the SPR Microarray for Detection of Insulin and HbA1c in 20-Times Diluted Whole Blooda
stepwise assembly on the Au chip | % change in reflectivity | |
---|---|---|
| ||
Insulin | HbA1c | |
Au chip-MPA-dendrimer | initial | initial |
Au chip-MPA-dendrimer-glutaraldehyde | 8.1 (±1.3) | 5.8 (±0.8) |
Au chip-MPA-dendrimer-glutaraldehyde-Ab | 7.4 (±1.4) | 7.7 (±1.0) |
Au chip-MPA-dendrimer-glutaraldehyde-Ab-0.1% | 0.9 (±0.5) | 1.0 (±0.4) |
BSA |
N = 3 replicates with eight spots per sample.