Abstract
We present a new class of carbon-based neural probes that consist of homogeneous glassy carbon (GC) microelectrodes, interconnects and bump pads. These electrodes have purely capacitive behavior with exceptionally high charge storage capacity (CSC) and are capable of sustaining more than 3.5 billion cycles of bi-phasic pulses at charge density of 0.25 mC/cm2. These probes enable both high SNR (>16) electrical signal recording and remarkably high-resolution real-time neurotransmitter detection, on the same platform. Leveraging a new 2-step, double-sided pattern transfer method for GC structures, these probes allow extended long-term electrical stimulation with no electrode material corrosion. Cross-section characterization through FIB and SEM imaging demonstrate strong attachment enabled by hydroxyl and carbonyl covalent bonds between GC microstructures and top insulating and bottom substrate layers. Extensive in-vivo and in-vitro tests confirmed: (i) high SNR (>16) recordings, (ii) highest reported CSC for non-coated neural probe (61.4 ± 6.9 mC/cm2), (iii) high-resolution dopamine detection (10 nM level - one of the lowest reported so far), (iv) recording of both electrical and electrochemical signals, and (v) no failure after 3.5 billion cycles of pulses. Therefore, these probes offer a compelling multi-modal platform for long-term applications of neural probe technology in both experimental and clinical neuroscience.
Introduction
Carbon is increasingly finding wider acceptance in micro and nanofabrication of a variety of sensors, actuators, microelectrodes, wires, batteries, fuel cells, thin-films, and neural interfaces1–6. Its unique tunable mechanical and electronic properties enabled by the availability of a range of possible hybridized bonds (sp2 and sp3) make it a versatile material. Further, carbon has a potential to assume even more far-reaching importance in applications varying from sensors to energy conversion and storage with discovery of newer carbon allotropes such as graphene (excellent conductivity and strength)7, Q-carbon (excellent hardness)8 and compressed glassy carbon (excellent strength)9. Specifically, in recent developments, the ability to microfabricate photolithographically patterned carbon features has further driven innovation in wider application of GC, one of the allotropes of carbon1,2. This technology involves pyrolyzing pre-patterned polymeric structure made from negative tone resists at high temperatures and inert environment in a tube furnace1,2.
Carbon is a compelling material for chronically implanted devices aiming to minimize tissue damage due to its chemical inertness, biocompatibility, good electrical properties, electrochemical stability, purely capacitive charge injection (no irreversible reactions and byproducts), and fast surface electrochemical kinetics1–6. We recently demonstrated neural probes with GC microelectrodes for electrocorticography (ECoG), intracortical recording and stimulation3–5. Brain stimulation applications require high performance microelectrode materials capable of sustaining hundreds of millions of cycles of electrical stimulations pulses (corresponding to implantations lasting 3–5 years) at clinically relevant charge densities without corrosion, tissue damage, or delamination of insulating and substrate layers. Meeting this requirement will enable the transition of neural probe technology from lab to clinical use, particularly with applications requiring long-term electrical stimulation, such as deep brain stimulation or spinal stimulation used for neuromodulation and induction of neuroplasticity10.
As a response to this need, we introduce here a new class of GC neural probes exclusively fabricated from a single homogeneous material of GC with no intermediate layers such as adhesion and conducting metal layers. This process removes any sources of potential mechanical or electrical discontinuities or failures at large cycles of electrical stimulations and extended period of in-vivo use. The new fabrication method presented here allows for a 2-step double-sided patterning, subsequent transfer to a polymeric supporting layer and final insulation of GC microelectrodes, interconnects and bump pads using a single material. In this new approach, therefore, there is no intermediate metal deposition process involved allowing–for the first time–all GC neural probes (aGC neural probes) made of a single homogenous material and supported on flexible polymeric substrate. The final aGC neural probes are housed in a flexible, thin-film polyimide substrate that leverage the superior electrochemical stability of GC material along with the improved biocompatibility of thin-film devices. To establish the robustness and superiority of the technology, extensive in-vivo and in-vitro mechanical, electrical, and electrochemical characterizations coupled with aggressive long-term stability and corrosion tests under electrical stimulations were carried out. This study demonstrates robustness and multi-modal recording capability for both neuronal electrical signals and neurotransmitter electrochemical signals. These probes are also capable of delivering balanced-charge pulsing over 3.5 billion cycles in extended accelerated aging process lasting more than 1000 hours. We demonstrate that these new probes have compelling advantages over existing technologies in almost all key metrics from mechanical and electrical properties to electrochemical kinetics, which is quite remarkable for probes fabricated from a single material without any surface coating.
Materials and Methods
Microfabrication
To demonstrate the technology, we microfabricated a 15-channel neural probe specifically designed for electrocorticography (ECoG) reading and a 4-channel penetrating probe designed and sized for neurotransmitter detection. The geometry of the ECoG device consisted of 300 μm diameter surface microelectrodes with 180 μm wide interconnects and a total width of 3.5 mm2,3. The neurotransmitter detection version had 4 penetrating microelectrodes of exposed area of 25 μm × 20 μm (500 μm2) spaced at 220 μm (see Supplementary Figures S1 and S2).
As shown in Fig. 1, the microfabrication process started with spin-coating 6 μm thick SU8–10 negative photoresist (Microchem, MA) and patterning it following conventional negative lithography process steps1,2. The patterned SU-8 microstructures were then pyrolyzed in a quartz furnace (MTI Corp., CA) under inert atmosphere of nitrogen gas with flow rate of 50 ml/min. For pyrolysis, the temperature was ramped to 1000 °C over a period of 6 hours and then held at this temperature for 90 minutes. Following this, the furnace was allowed to cool down to room temperature over a period of several hours.
Subsequently, the first base layer of ~8 μm thick non-photo sensitive polyimide (Durimide 115a) (Fujifilm, Japan) was spin-coated on top of GC microstructures at 2500 rpm for 45 seconds. This was then soft baked at 135 °C for 2.5 min and cooled down to room temperature. Patterning of Durimide was done through dry reactive ion etching (Technics Series 85 Reactive Ion Etcher, Pleasanton, CA) using Shipley positive resist (Microchem, MA) as a soft mask. The patterned Durimide layer was partially cured at 200 °C for 30 mins in N2 environment. A sacrificial transfer layer was then deposited as a secondary substrate on top of the Durimide layer. Polymers such as PMMA (Polymethyl methacrylate) or PDMS (Polydimethylsiloxane) could be used for this purpose. Here, we describe the use of PDMS (Dow Corning, USA) as one example. PDMS mixed with curing agent in 10:1 proportion was spin-coated at 300 rpm for 45 secs and then cured at 90 °C for 30 mins followed by BHF etching of oxide layer to release it from the silicon substrate. Subsequently, a glass substrate was bonded to the released structure to offer an additional temporary support for improved alignment. The wafer was then flipped and an insulation layer of Durimide spin-coated. After patterning this top insulation layer of Durimide, it was subsequently partially cured under N2. A final BHF etching removed the sacrificial layer and released all devices. The final aGC probes are shown in Supplementary Figures S1 (μECoG probe) and S2 (intracortical penetrating probe).
Mechanical and Cross-Sectional Characterizations
The mechanical strength of the ensuing device was measured using Instron’s 1500HDX Universal Testing Machine (Instron, USA) where tensile load was applied till failure, and the resulting extension measured. Tensile test specimen probes with dimensions of 17 mm in length, 0.5 mm in width and 50 μm in thickness were used for the experimental study. Since the devices were too thin, attachment of strain gauge extensometers directly to the probes was not practical. Instead, linear displacement due to the crosshead motion of Instron was used to obtain displacement values. Thread-rolled C-clamps with clamping capacity of 0–25 mm were used to grip the probe. C-clamps provide greater precision for easy adjustment of position and prevent marring of the probe surface. The frame of C-clamps was then held between the jaws of upper and lower clamps of the Instron machine (ca. Supplementary Figure S4). The cross-head extension rate was set at 0.5 mm/min. The specimen was then loaded to failure at this rate and the load-deflection curve plotted. For comparison purposes, probes of similar geometry but consisting of 200 nm thick metal (Pt) microelectrodes and wire traces were microfabricated through standard metal lift-off process. Subsequently, the load-deflection curves and ultimate load carrying capacity of the thin-film metal probes were determined in a similar fashion. Young’s Modulus was determined by taking the mean slope of ten points in the elastic region of the stress-strain curve. The modulus of plain Durimide substrate with no metal or GC interconnects and microelectrodes was also determined separately to de-couple its effect from the composite structure. FEI Quanta 450 FEG SEM (ThermoScientific, OR) was used for scanning electron microscopy images. Cross sectioning was obtained through Focused Ion Beam (FIB) milling using a dual beam microscope system (Nova 600 NanoLab; FEI, Netherlands). This system is equipped with a field emission gun for scanning electron imaging and a focused ion beam of gallium ions for milling. The sectioning process was done in two steps: ion current milling (working at 30 kV and 7.7 pA) followed by a cleaning step of the section. Imaging was done at an acceleration voltage of 20 kV (0.4 nA of emission current).
Electrical, Electrochemical, and Stability Characterizations
The electrochemical behavior of the microelectrodes was studied in Phosphate-Buffered Saline solution (PBS; 0.01 M, pH 7.4; Sigma Aldrich, USA). Cyclic voltammetry (CV) was used to quantify capacitive charging and microelectrode stability while electrochemical impedance spectroscopy (EIS) was used to determine the electrical properties of the probes over a large range of frequencies. For both CV and EIS tests, we used a potentiostat (Reference 600+, Gamry Instruments, USA) connected to a three-electrode electrochemical cell with a platinum wire as a counter electrode and a saturated Ag/AgCl reference electrode. During CV tests, the working electrode potential was swept between 1.3 V and −0.9 V (water window of GC4), maintaining a scan rate of 100 mV/s. The total charge storage capacity (CSC) was calculated from the time integral of a CV cycle. During EIS measurements, 10 mV RMS amplitude sine wave was superimposed on 0 V potential with frequency sweep from 0.1 to 105 Hz. Equivalent circuit modeling of the EIS data was done through Gamry Echem Analyst Vn 7.05 software (Gamry Instruments, USA). For the accelerated aging test consisting of prolonged current stimulation patterns, we used a cathodic-first charge balanced bi-phasic current pulses with 450 µA amplitude, 400 μsec cathodic half-phase period and a frequency of 1 kHz in saline solution (0.01 M PBS) using a Gamry Instruments Virtual Front Panel (Gamry Instruments, PA, USA). This accelerated aging test corresponds to charge density of 0.25 mC/cm2 and was applied over a period of 1000 hours. Voltage transient responses during stimulation were acquired using the same potentiostat by simultaneously injecting stimulation current pulses and recording the corresponding voltage excursions between working and counter electrodes. In regular intervals, impedance was measured and the integrity of the electrodes checked. The buffer pH was measured regularly during accelerated aging test using pH/mv/ion benchtop meter (Jenco Electronics, TX).
Fast-Scan Cyclic Voltammetry (FSCV) was used to evaluate the electrodes’ dopamine detection performance with a WaveNeuro Potentiostat System (Pine Research, NC). For in-vitro dopamine calibration, the FSCV waveform was used with a pyramidal excursion from −0.5 V to +1.3 V potential and back to baseline with respect to the Ag/AgCl reference electrode at a scan rate of 400 V/s and 10 Hz frequency11. The duration of each scan was 9 ms (900 data points). Prior to the beginning of each experiment, the same voltage waveform was applied to the microelectrodes at 60 Hz for 1 hour for activating the carbon surface of the microelectrodes. Known concentrations of dopamine (10 nM–1 μM) were then infused over 5 seconds while changes in current were recorded for 20 seconds.
In-vivo Tests
Micro-ECoG Implantation
All animal experiments were performed in accordance with the Association for Assessment and Accreditation of Laboratory Animal Care (AAALAC) Guide for the Care and Use of Laboratory Animals (8th Edition) and approved by the University of Washington Institutional Animal Care and Use Committee (IACUC) under protocol number 4265-01. Adult female Long-Evans rats (300 g) were used in this study and anesthesia was induced with a mixture of ketamine and xylazine. The animal was placed in a stereotaxic frame and a craniotomy made 1.5 mm rostral and 4.5 mm caudal to the bregma and 1 mm lateral and 7 mm lateral to the midline, targeting the left forelimb sensorimotor cortex. The microelectrode array was then placed on the dura and stabilized by gel foam dental cement. A reference wire was connected to a screw on the right parietal bone while a ground wire was connected to a screw on the occipital bone. Throughout the entire procedure, body temperature of the animal was maintained with a heating pad placed under it.
Electrophysiology Recording
Brain signal recording sessions were carried out under anesthesia, 4 days after array implantation. The implanted ECoG microarray was connected to a multichannel data acquisition system (Tucker-Davis Technologies, FL) through an active (unity-gain) head-stage. Sensory evoked potentials (SEP) were obtained by applying bipolar surface stimulation to the surface of skin of the right wrist of an anesthetized rat. Recorded data was digitized at 24.4 kHz and stored on a PC for further analysis. Subsequently, raw signals were bandpass filtered (1–450 Hz, 4th order Butterworth, zero-phase) and down-sampled to 1 kHz using Matlab software (Mathworks, Inc., MA). Low-frequency brain activity in evoked SEPs were eliminated by a high-pass filter (30 Hz, 4th order Butterworth, zero-phase). In each recording session, 15 single, bi-phasic, squared wave stimulation pulses (pulse width: 500 µs, pulse amplitude: 0.5 mA − 2 mA, 0.5 Hz) were applied to the surface electrodes and evoked response from 5 ms to 55 ms after stimulation onset were averaged to obtain average SEPs12. The extracted evoked response between 5 ms to 55 ms after stimulation onset was considered as an estimate for the signal, while the extracted spontaneous activity corresponding to 65 ms to 115 ms after stimulation onset was taken as an estimate for the background noise. Subsequently, SNR (signal-to-noise ratio) of ECoG electrodes placed in six different positions across forelimb sensory cortex were evaluated. The SNR value for each trial was defined as the ratio between the variance of selected signal (evoked response) and the variance of selected noise (spontaneous activity). The calculated SNR (mean ± standard deviation) of all channels was used as a useful indication of the quality of brain-signal recording. Furthermore, to show temporal-spectral representation of recorded data by GC microelectrodes, the spectrogram of each ECoG channel were computed by applying 256 ms Hanning window on each ECoG time-series. The corresponding short-time Fourier transform of each segment were then computed using 200 ms overlap.
Results
In this section, we present the outcomes of the mechanical, electrical, and electrochemical characterizations.
Optical and Cross-Sectional Characterizations
Optical microscopy and SEM were used to visually investigate the structural integrity of the probes. As shown in Fig. 2a and Supplementary Figures S1–S3, the microelectrodes and interconnects (wire traces and bump pads) look sharp and well-defined after a two-stage, double-sided, pattern transfer procedure. Both the top and bottom insulation layers appear structurally robust. As can be seen in Fig. 2b,d, FIB characterizations of GC microelectrode/substrate and GC interconnect/substrate interfaces demonstrate that the top insulation and bottom substrate polymer layers have an almost seamless interface with GC suggesting a strong bond between GC and polymer layers. This smooth and seamless interface is an important requirement for a tight integration and composite action between GC and the polymer layers.
Mechanical Characterizations
Results from mechanical characterization are summarized in Fig. 3 showing failure planes at maximum load carrying capacity. Load-deflection curves under a tensile load are reported along with calculated modulus (Supplementary Figure S4). For comparison purposes, tensile load tests were carried out on probes with thin-film metal (Pt) microelectrodes and interconnects with similar geometry as their aGC counterparts. The load carrying capacity for both types of probes was directly obtained from the load-deflection curves and the resulting failure planes in transverse direction were further investigated using SEM. The aGC probes failed at a tensile load of ~12 N at 1 mm extension with clearly defined elastic and plastic regions observed in the load-deflection curves. On the other hand, probes microfabricated with thin-film metal (Pt) microelectrodes and traces failed at an ultimate tensile load of 5 N at 0.7 mm extension. For both thin-film metal and GC traces, the failure plane as shown in Fig. 3a,b passes through the traces with no jagged discontinuity, suggesting a strong bond between durimide and the respective metal or GC layer. The Young’s Modulus of the composite structure with GC microelectrodes and traces was calculated to be ~2.65 GPa whereas it was ~2.59 GPa for probes with Pt microelectrodes and traces. The modulus of aGC probes was slightly higher than that of plain Durimide (2.5 GPa) suggesting a strong composite action between GC and polyimide3. Optical observation revealed several micro-cracks in the thin-film metal traces near the failure plane (Fig. 3a), whereas only a single additional crack was seen in the otherwise intact GC trace (Fig. 3b), indicating different failure modes in the two types of probes. The extent of micro cracks in the thin-film metal electrodes suggests that they were progressively produced during loading resulting in earlier onset of electrical connection failure. For aGC probes, failure in the traces under tensile load seems to have occurred at higher load levels preceding plastic failure. Stiffness of aGC probes in compression was determined through the Euler buckling load: Pcr = π2EI/(KL)2, where E = Young’s modulus, I = area moment of inertia (500 μm width and 50 μm depth), K = effective length factor (usually 1) and L = length of probe. For the intracortical penetrating probes reported here, Pcr = 2.13 mN, which is substantially higher than ~1.5 mN force required to penetrate the cerebral cortex of a rat13.
Electrical, Electrochemical and Stability Characterizations
Electrical characterization results are shown in Fig. 4. The impedance values at 10 Hz, 100 Hz and 1 kHz are (42.5 ± 27.6) kΩ, (10.7 ± 4.9) kΩ and (5.8 ± 1.2) kΩ, respectively (mean and standard deviation, n = 10). These indicate an excellent range for micro-ECoG recording applications. The corresponding EIS plots are reported in Fig. 4a. As the phase plot in Fig. 4a shows, impedances at frequencies higher than 4 kHz are dominated by almost exclusively a resistive component. Results from cyclic voltammetry of GC microelectrodes are summarized in Fig. 4b demonstrating an approximately rectangular shape, characteristic of microelectrodes exhibiting a predominantly double-layer capacitance. This is consistent with several other reports showing that this type of behavior is exhibited by only purely carbon materials, indicating that no faradaic reactions occur during the charging and discharging process4,14–17. CSC, calculated as the time-integral of an entire CV cycle between the water oxidation and reduction potential limits, is 61.4 ± 6.9 mC/cm2 (mean and standard deviation, n = 10).
This capacitive charge injection mechanism is also confirmed by the voltage transient shape (Fig. 4c) illustrating the comparison of voltage transients between GC and Pt microelectrodes with the same dimension of 300 µm diameter and comparable impedance. Traces are in response to a biphasic, symmetric current pulse of 450 µA amplitude and 400 μsec of half-phase period. Due to the capacitive charge injection mechanism of GC microelectrodes, the amplitude of voltage transient for the same current level applied to both types of probes is much less in GC microelectrodes, producing maximum negative voltage (Vmax,neg) of −1.03 V and maximum positive voltage (Vmax,pos) of 0.25 V. These correspond to Emc (most negative polarization voltage) across electrode-electrolyte interface of −0.75 V and Ema (most positive polarization voltage) close to 0 V18. These values are safely within the maximum potential limits of water reduction and oxidation for GC microelectrodes (i.e., −0.9 to 1.3 V; Fig. 4b). In contrast, for Pt microelectrodes, the corresponding Vmax,neg, Emc, Vmax,pos and Ema values are −2.25 V, −1.75 V, 1.07 V and 0.7 V, respectively. These values are substantially outside the maximum potentials corresponding to the water window of Pt at −0.6 V to 0.8 V. Note that the maximum negative potential excursion (Emc) is calculated by subtracting the access voltage (Va) associated with the Ohmic electrolyte resistance and overpotential terms from the maximum negative voltage in the transient (i.e., Emc = Vmax,neg − Va and Ema = Vmax,pos − Va)18. Va is defined as the near-instantaneous voltage change at either the onset of the current pulse or immediately after the current pulse is terminated18. In this condition, Va is 0.36 V for GC and 0.48 V for Pt (33% higher for Pt), suggesting a substantially lower overpotential to be overcome in GC microelectrodes (Fig. 4c,d). Table 1 summarizes these results.
Table 1.
Material | ΔV | Eipp | Va | ΔEp | Emc | Ema | Vmax,neg | Vmax,pos |
---|---|---|---|---|---|---|---|---|
GC | 0.72 | −0.36 | 0.36 | 0.36 | −0.75 | 0.05 | −1.03 | 0.25 |
Pt | 1.88 | −0.31 | 0.48 | 1.4 | −1.75 | 0.71 | −2.25 | 1.07 |
Emc = (Vmax,neg − Va), Ema = (Vmax,pos − Va), and ΔEp = ΔV − Va.
Further, to gain better understanding of the charge transfer properties of the microelectrodes, an equivalent circuit model was curve-fitted to the experimental EIS data. Representative EIS bode plots of the experimental data and curve-fitting of the most appropriate equivalent circuit model are reported in Fig. 5. The optimal model with the best fit is a modified Randles circuit that is composed of electrolyte resistance (Rsol) in series with a parallel combination of charge transfer resistance (Rct), constant phase element (ZCPE), and Warburg impedance (Zw)19–21. The modified Randles equivalent circle selected here is consistent with results reported in the literature for capacitive microelectrodes18–21. In this model, the constant phase element ZCPE is define as: ZCPE = 1/(jω)αYo where Yo represents the capacitance and α is a constant related to the angle of rotation in the complex plane compared to the purely capacitive behavior (Yo = 1 for a pure capacitor)19,20. The corresponding parameters obtained by fitting the experimental data to the model are: Rsol = 4.4 kΩ, ZCPE with Y0 = 4.07e−11 S*sα and α = 0.87, ZW = 2.74−11 S*s1/2, and Rct = 3.34 kΩ.
Results from accelerated aging test with cycles of electrical pulse applied over period of 1000 hours are reported in Fig. 6a,b. Figure 6 summarizes EIS plots, CV cycles and voltage transients of GC microelectrodes before and after 0.2, 1, 2, 2.5 and 3.5 billion cycles of electrical stimulation pulses. Prior to pulsing, for example, the impedance value at 1 kHz was ~5.2 kΩ. After 2, 2.5 and 3.5 billion cycles of pulses, the impedance at 1 kHz was measured to be 5.75, 6.3, 8.8 kΩ, respectively. In general, we observed a decrease in impedance at low frequency ranges with a shift of the low frequency pole and a corresponding increase in CSC. The CSC values increased from 45 mC/cm2 to 180 mC/cm2 after 0.2–3.5 billion cycles of pulsing. This increase in capacitance is predominantly due to continued surface activation that is very common in carbon microelectrodes22. The corresponding CV cycles are reported in Fig. 6b, exhibiting the expected rectangular shape of capacitive microelectrodes. Again, as CV diagrams demonstrate, there are no observable faradaic reactions occurring during charging and discharging processes.
As part of this accelerated aging test, pH value of ~7.55 was measured at several locations in the flow-cell, well within the buffer’s initial pH of 7.4. This is expected due to the buffering action of the PBS solution. Moreover, our findings are consistent with pH levels predicted by a transient FEA model in our previous work where pH near the microelectrode surface returned back to the initial conditions after each charge-balanced cycle23.
SEM and AFM inspections of GC microelectrodes demonstrate that the mechanical integrity of the microelectrode was still intact after 3.5 billion cycles of biphasic 0.25 mC/cm2 pulses in PBS solution over a period of 42 days (Fig. 7). SEM images confirm that the insulation and substrate polymeric layers remained intact and well attached to GC microelectrodes at the end of this accelerated aging under electrical stimulation. Further, as shown in Fig. 7, AFM images of the microelectrodes were taken at 25 μm × 25 μm area before and after aging to determine if there was any quantifiable and visible corrosion of the exposed microelectrodes. AFM morphology analysis indicated that, before stimulation, mean roughness of the microelectrode surface was 43.55 ± 3.09 nm (n = 4) while after stimulation, it was 44.00 ± 6.69 nm (n = 4). A corresponding t-test gave a nonsignificant difference (P value > 0.05) between the mean roughness of the surfaces under the two conditions, confirming that there was no discernible corrosion after the aging process.
To test GC microelectrodes at near their charge injection limit, we applied a cathodic-first charge balanced bi-phasic current pulses with 5.1 mA amplitude, 400 μsec cathodic half-phase period and a frequency of 1 kHz in saline solution (0.01 M PBS) for 50 million cycles of pulses (3 mC/cm2). Voltage transients of a GC microelectrode in response to this biphasic pulse together with impedance spectra and the corresponding CV plots before and after 20, 40, 50 billion stimulation pulses are reported in Supplementary Figure S5, demonstrating that the microelectrode was able to sustain aggressive stimulation without significant changes in electrochemical performances.
In-Vivo Electrophysiology Tests
SEPs caused by the bi-phasic stimulation pulses on the right wrist were recorded via the ECoG microelectrode array implanted on the left forelimb sensory cortex of an anesthetized rat. Figure 8a shows the craniotomy made targeting the left forelimb sensorimotor cortex where the probes were implanted. Plots of brain activities sampled from six channels of GC microelectrodes implanted in the sensory cortex four days after the implantation surgery are shown in Fig. 8b. To verify the quality of brain recording using GC microelectrodes, the average-SEP of different channels were calculated in response to the surface electrical stimulation of hand area. Figure 8c shows a typical average-SEP (mean ± standard error) of a single ECoG channel for 5 ms to 55 ms after stimulation onset, with average-spontaneous activity of the same ECoG channel for 65 ms to 115 ms after stimulation onset. The time frame between 5 ms to 55 ms was considered adequate to cover the entire SEP response, as all SEPs were observed in this time range. The large positive and negative response corresponding to 15 ms after stimulation onset demonstrates that the ECoG microelectrodes successfully recorded early response of sensory neurons in somatosensory cortex in the presence of background spontaneous activity. Figure 8d shows the power spectrum of the same six channels. As can be seen, all of the microelectrodes recorded brain activities in physiological spectral range without interference from non-physiological noise such as power line noise. Further, Supplementary Figure S6 shows that the spectrograms representing brain signals could be recorded across different frequency bands using GC microelectrodes. In Table 2, the mean (±standard deviation) SNR value (16 ± 6) and impedance of all ECoG channels are shown. Collectively, the in-vivo tests demonstrate that the GC ECoG microarray can successfully record multi-site, temporal and spectral brain information with high SNR.
Table 2.
SNR (mean ± std) | In-vivo Impedance of Microelectrodes (kΩ) (mean ± std) (n = 8) | Impedance of Ground Electrode (kΩ) |
---|---|---|
16 ± 6 | 81 ± 16 | <0.1 |
Neurotransmitter (Dopamine) Detection
Figure 9 summarizes the results obtained from a series of FSCV characterizations on a 500 μm2 net surface area penetrating GC microelectrode probes (Supplementary Figure S2). Figure 9a shows an example of dopamine signals detected using a standard pyramidal FSCV waveform in which the applied voltage was ramped from the holding potential of −0.5 V to the switching potential of +1.3 V and then back to −0.5 V at 400 V/s scan-rate. In particular, the color plots show changes in current following the injection of 10 nM of dopamine (DA) at t = 15 secs first and subsequently an additional 10 nM at t = 25 sec in the base PBS buffer. These changes in currents occurred at the known reduction and oxidation peaks of DA (i.e., −0.18 V and 0.62 V), as clearly shown in the background subtracted CVs for 10 nM and 20 nM DA given in Fig. 9b. Further, to determine a calibration curve over a wider range of concentrations, additional sets of FSCV were carried out at 10,20,50,100,200,500,700 nM and 1 μM concentration of dopamine in PBS. As shown in Fig. 9c, the background subtracted CVs corresponding to these range of concentrations demonstrate reduction and oxidation peaks at (−0.18 V) and (0.62 V). In addition, the calibration curve for these concentrations of DA varying from 10 nM to 1 μM follows a linear trend as reported in Fig. 9d. This low detection limit of dopamine at 10 nM represents one of the lowest reported limits in the literature.
Further, the capability of this aGC platform for selective detection of dopamine in the presence of Ascorbic Acid (AA) was also investigated. AA typically occurs in much higher concentration than that of DA (100–1000 times) and strongly interferes with DA detection24. Hence, the determination of DA with high selectivity and sensitivity in presence of AA is desirable for diagnostic applications. GC microelectrodes were able to detect 50 nM DA in the presence of 1 mM AA with high resolution. Background subtracted CVs showing reduction (−0.2 V) and oxidation (0.6 V) peaks of 50 nM and 500 nM of DA in presence of 1 mM AA in PBS are reported in Fig. 9e. The corresponding calibration curve shows a linear trend (Fig. 9f). This capability uniquely underscores an additional strong asset of GC microelectrodes in neurotransmitter detection through FSCV, a capability that is not matched by thin-film metal or metal oxide microelectrodes.
Discussions
This study aims to overcome key barriers towards long-term clinical translation of neural probe technology. We demonstrate (i) long-term stimulation of electrodes at biologically relevant charge densities, (ii) ability to electrochemically sense neurotransmitters with the same platform as electrical signal recording, and hence potentially enable understanding of the interplay between the two signaling modalities, (iii) no corrosion of the electrode material and, (iv) no delamination of layer. The extensive characterizations reported here are targeted towards investigating these capabilities and establishing the suitability and superiority of GC microelectrodes towards advancing clinical translation of neural interfaces. Below, we review these in light of key results reported in the literature for other competing probe materials.
The mechanical and electrical integrity of the aGC probes, when exposed to aggressively large number of electrical stimulations cycles, is documented via SEM, AFM, FIB, EIS and CV characterizations. AFM and SEM images taken at several locations on the microelectrodes after accelerated aging show that there is no measurable and statistically significant difference on the surface roughness of the microelectrodes before and after aging. This strongly supports the observation that GC experiences no corrosion for voltage excursions applied within its wide electrochemical window of 2.4 volts4. Further, the maintenance of strong mechanical integrity of the probes after 1000 hours of accelerated aging under active electrical stimulation indicate that the attachment between GC microstructures and polyimide remains excellent. We had previously shown that the interface between GC and polyimide is dominated by strong hydrogen bonds through hydroxyl and carbonyl groups as well as covalent bonds through anhydride groups25.
It is important to closely examine if this exceptional performance in mechanical integrity is accompanied with equally superior electrical and electrochemical performance. For that, a closer examination of surface electrochemical kinetics in GC microelectrodes will provide useful parameters for assessing the performance of these microelectrodes under recording and electrical stimulation conditions. For example, CSC is typically an excellent indicator of the type of dominant current transduction in a microelectrode (i.e., faradaic Vs double-layer charging). In this study, we show that for GC microelectrodes made of a single and homogenous material with no coating of any kind, CSC is as high as 61.4 ± 6.9 mC/cm2. This value is slightly lower than that of carbon nanotubes (CNT) directly grown on metal microelectrodes (70.8 ± 1.1 mC/cm2)14,15, but much larger than porous graphene (50 mC/cm2)17. For further comparison, CSC for PEDOT-PSS coatings is 40.4 ± 4.8 mC/cm2 while it is 893.5 ± 137.8 mC/cm2 for PEDOT-PSS-CNT coatings and ~90 mC/cm2 for AIROF (Activated Iridium Oxide Film) microelectrodes26–30. All these microelectrode materials, however, utilize a coating on top of a thin-film metal layer. In the case of AIROF, it has an additional drawback of transferring charges through faradaic reactions that could result in microelectrode degradation and tissue damage. Therefore, our data support the interpretation that GC has the highest reported CSC for a single and homogenous probe material. In fact, if the surface of GC could be made porous through plasma etching or similar processes, the CSC value may further increase significantly.
Additional insight into electrode kinetics in GC microelectrodes can be obtained from voltage transients. The voltage transients in GC, in both negative and positive excursions, show a rapid and almost instantaneous convergence in the voltages suggesting a very fast electrode polarization. Pt microelectrodes, on the other hand, show a steep slope and slower polarization before Vmax-neg or Vmax-pos are reached. The voltage residuals at the end of each pulse cycle are also revealing in terms of the speed of charging and discharging. GC microelectrodes almost instantaneously converge back to starting potential whereas Pt still has significant voltage residual. In addition, electrode overpotential (η), which represents the deviation from the equilibrium potential (E − Eeq), is mainly driven by redox process and is a very good indicator of surface electrode kinetics18,29–31. The net voltage drop at the electrode is a sum of the Ohmic IR drop, electrode overpotential (drives redox at electrode/electrolyte interface), solution overpotential (due to transient charge distribution), and the shift in the equilibrium potential of the electrode (i.e., ΔV = IR + ηelectrode + ηsol + ΔEo)18. Here, the same buffer is used for GC and Pt microelectrodes. Therefore, the differences in ΔV between the two materials are mostly due to electrode overpotential (ηelectrode), and to a lesser extent on solution overpotential (ηsol). The overpotential of GC is significantly lower than that of Pt (0.38 Vs 0.48 V) suggesting that the voltage drop at the GC microelectrode surface due to overpotential accompanied by double layer charging and discharging is low (i.e., no reaction occurs). This is expected as overpotential is mainly influenced by faradaic reactions that are dominant in Pt microelectrodes.
With regard to accelerated aging under electrical stimulation, the charge density used in the experiments reported here (i.e., 0.25 mC/cm2) is consistent with the charge density used to evoke sensations in humans using millimeter-scale ECoG electrodes32. In the literature, there have been several reports of accelerated aging tests with and without electrical stimulation. With respect to passive (non-electrical) accelerated aging, thin-film Pt microelectrodes with adhesion promoters (i.e., polyimide-DLC-SiC-Pt) subjected to constant temperature of 37 °C for a period of 1 year33, silicon carbide microelectrodes immersed in a buffer heated to 96 °C for 1600 hours34, and PEDOT-coated microelectrodes immersed in buffer for 4 weeks have been reported28. These tests, however, were limited to either mechanical shear strength or current leakage of the probes. Electrical stimulation tests were reported for iridium microelectrodes with 2 billion cycles applied over ~167 hours (0.25 mC/cm2)35, porous graphene microelectrodes under 1 million biphasic current pulses17, GC with adhesion promoters (SiC and DLC) under 20 million pulses (0.2 mC/cm2)36, CVD-CNT-coated electrodes under 1 million pulses (1.6 mC/cm2)15, and PEDOT-coated electrodes subjected to 24-hour biphasic stimulation (3 mC/cm2)29. In general, the use of coatings is often a point of weakness, especially in cases where there is active electrical stimulation. A variety of materials such as titanium nitride, iridium oxide, PEDOT, and carbon nanotube composites have been adopted as coating layers37–47. However, these materials are prone to delamination during stimulation and long-term implantations, caused typically by poor adhesion between the coating material and the metal electrode37,38,45. Our current work is the first reported case where any neural probe, either with or without a coating, is subjected to 1000 hours and 3.5 billion cycles of active electrical current pulses. During these tests our probe exhibited no corrosion, delamination, or any sort of mechanical or electrical failure. Furthermore, on the same GC microelectrodes, we carried out an even more aggressive test at charge injection limit of 3 mC/cm2 with a cathodic-first charge balanced bi-phasic current pulses consisting of 5.1 mA amplitude, 400 μsec cathodic half-phase period and a frequency of 1 kHz in saline solution (0.01 M PBS) applied for 50 million pulses. Once again, GC microelectrodes survived this extreme test with no delamination of polymeric layers and no corrosion of electrodes. Further comparisons are given in Supplementary Table 1.
What drives this resistance to corrosion in GC microelectrodes? In the case of GC microelectrodes reported here, the ability to withstand such a high number of cycles of pulses can be explained by the fact that (i) GC is a homogeneous material with no crystalline structure; hence devoid of granular surfaces that could act as points of initiation of corrosive chemical reactions during stimulation1,2 and (ii) weaker metal adhesion layers have been eliminated. A previous study using AFM and SEM characterization demonstrated that surface roughness of GC is only a few nanometers, an order of magnitude smaller than typical thin-film metal electrodes4,47. In addition to this smoothness in surface, the current study has demonstrated that absence of a thin-film adhesion (Cr or Ti) and thin-film metal layer (Pt or Au) eliminates any potential Galvanic corrosion in probes48. This is a key distinguishing aspect of the technology presented here.
In addition, our demonstration of high-resolution real-time in-vitro neurotransmitter detection through background-subtracted FSCV using GC microelectrodes highlights the versatility and multi-modal (electrical and electrochemical) recording capabilities of the platform. Driven by its high spatiotemporal resolution and ability to detect sub-second fluctuations in neurotransmitter concentrations, FSCV has become a technique of choice for neurotransmitter detection. The 10 nM limit of detection (LOD) for dopamine reported here is one of the lowest ever reported in the literature. Given that biologically relevant dopamine concentration characteristic of living systems is within 26–40 nM49, such a low LOD represents significant progress in this area. There are several reports in the literature that document similar level of sensitivity. For example, CNT directly grown on metal microelectrodes (CNT-Nb) have shown dopamine detection limit of 11 ± 1 nM50, whereas SWCNT (Single Walled CNTs) on metal microelectrodes have shown LOD of 17 nM51. Further, for carbon nanofibers modified by PEDOT/graphene oxide coating, LOD of 85 nM was reported52 while laser-treated CNT yarn microelectrodes (CNTYMEs) have been shown to have LOD down to 13 ± 2 nM53. Further comparisons are given in Supplementary Table 2. As shown in the table, the GC LOD of 10 nM is superior to all other probes measured. Further, it is noteworthy that the majority of FSCV experiments reported in the literature were performed using carbon-fiber microelectrodes or single microelectrodes, allowing only single measurements. Glassy carbon microelectrodes, on the other hand, are suited for multi-site FSCV detection, similar to reported 3D carbon nanofiber microelectrode arrays (MEAs) fabricated through O2 plasma-assisted pyrolysis54, and pyrolyzed carbon MEAs using thin positive resists on non-biocompatible silicon substrates that were used for dopamine detection55.
An obvious question is what drives this high neurotransmitter sensitivity? GC surfaces–as reported here –- are ideal for detecting neurotransmitter electrochemical activity due to the intrinsic hydroxyl groups that form on their surfaces25. These hydroxyl groups are favorable for adsorption of cationic species such as dopamine whose amine side chain gets protonated at physiological pH56. Further, the strength of adsorption of dopamine at GC surface (and hence sensitivity) has been shown to correlate with increased heterogeneous electron transfer kinetics. This is an advantage of GC, as demonstrated in the voltage transients of Fig. 4c–e.
Conclusions
In this study, we introduced the first use of all glassy carbon probe structure where microelectrodes, traces, and bump-pads are all fabricated from a single and homogeneous GC material. The probe experienced no measurable corrosion after a very aggressive accelerated aging test that involved 3.5 billion cycles of pulses. Through extensive characterizations, the versatility of this material, that may accelerate the translation of neuroprosthetic devices to clinical use, was demonstrated.
The following points highlight some of the key findings reported in this work:
The process introduced is compatible with MEMS and semi-conductor processes making the fabrication of these electrodes and their integration with CMOS a possibility.
GC microelectrodes have high SNR (>16) in-vivo recordings capability.
The accelerated aging introduced here is a record-setting 3.5 billion cycles of bi-phasic charge-balanced pulse train over period of 1000 hours. To the best of our knowledge, this is the first report on pulsing at this level and duration.
The corrosion rate in GC is very minimal and there was no difference between new and aggressively aged microelectrodes.
There was no failure at the GC and polyimide interface after 3.5 billion pulses and 1000 hours of immersion in a buffer. This suggests a strong mechanical robustness at the interface of microelectrodes and the supporting as well as insulating substrates aided by both hydrogen bonds (hydroxyl and carboxyl groups) and covalent bonds (anhydride groups).
GC microelectrodes have one of the highest reported CSC for single material non-coated neural probe (i.e., 61.4 ± 6.9 mC/cm2).
GC microelectrodes have one of the lowest limits of detection for dopamine at 10 nM. This is a significant finding.
The same probe materials are capable of stimulating, as well as recording both electrical and electrochemical signals (neurotransmitters). Therefore, GC microelectrodes offer a multi-modal platform that may accelerate investigation into the interplay between electrical and electrochemical signaling in the brain.
Collectively, our results demonstrate that GC may be an ideal microelectrode material for neuroprosthetics devices given its good impedance, excellent electrochemical response, high fidelity recordings, strong resistance to corrosion after billions of pulses of electrical stimulations, and ability to electrochemically detect neurotransmitters at one of the lowest concentrations reported. It is rare to find all these attributes in a single platform.
Electronic supplementary material
Acknowledgements
This material is based on research work supported by the Center for Sensorimotor Neural Engineering (CSNE), a National Science Foundation Engineering Research Center (EEC-1028725). We would like to thank Ms. Nha Uyen Huynh of MEMS Research Lab and Dr. George Yusuf of SDSU for AFM support and Dr. Ingrid Niesman, Director of SDSU EM Facility for SEM support.
Author Contributions
S.N. implemented the all-glassy carbon process, designed and fabricated the devices, performed electrochemical measurements, SEM imaging, electrical stimulation tests, helped with dopamine calibration tests, and contributed to writing the paper; E.C. designed and performed dopamine calibration tests, supervised and performed electrochemical and long-term stability characterizations, made electrochemical modeling, designed and performed voltage transient tests, and contributed to writing the paper (dopamine detection and electrical/electrochemical characterizations, and results sections); A.B. performed mechanical measurements; A.S. performed FIB characterizations; S.S. performed in-vivo neural recording experiments and contributed to writing section on in-vivo data; A.K. performed in-vivo neural recording experiments and contributed to writing section on in-vivo data; A.B. performed in-vivo neural recording experiments and contributed to writing section on in-vivo data; S.T. supervised neural recording experiments and helped with dopamine calibration tests; C.M. designed and supervised in-vivo tests and revised the paper; S.K. formulated all glassy carbon (aGC) probe concept and supervised the project, structured the outline of the paper and wrote discussion section of the paper.
Competing Interests
The authors declare no competing interests.
Footnotes
Surabhi Nimbalkar and Elisa Castagnola contributed equally to this work.
Electronic supplementary material
Supplementary information accompanies this paper at 10.1038/s41598-018-25198-x.
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Schueller OJA, Brittain ST, Marzolin C, Whitesides GM. Fabrication and Characterization of Glassy Carbon MEMS. Chem. Mater. 1997;9:1399–1406. doi: 10.1021/cm960639v. [DOI] [Google Scholar]
- 2.Ranganathan S, McCreery RL, Majji SM, Madou M. Photoresist-Derived Carbon for Microelectromechanical Systems and Electrochemical Applications. J. Electrochem. Soc. 2000;147(1):277–282. doi: 10.1149/1.1393188. [DOI] [Google Scholar]
- 3.Vomero M, et al. Novel Pattern Transfer Technique for Mounting Glassy Carbon Microelectrodes on Polymeric Flexible Substrates. Journal of Micromechanics and Microengineering. 2016;26:36–44. doi: 10.1088/0960-1317/26/2/025018. [DOI] [Google Scholar]
- 4.Vomero M, et al. Highly Stable Glassy Carbon Interfaces for Long-Term Neural Stimulation and Low-Noise Recording of Brain Activity. Nature Sci. Rep. 2017;7:40332. doi: 10.1038/srep40332. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Kassegne S, et al. Electrical impedance, electrochemistry, mechanical stiffness, and hardness tunability in glassy carbon MEMS µECoG electrodes. Microelectronic Eng. 2015;133:36–44. doi: 10.1016/j.mee.2014.11.013. [DOI] [Google Scholar]
- 6.Kozai TD, Kipke DR. Insertion shuttle with carboxyl terminated self-assembled monolayer coatings for implanting flexible polymer neural probes in the brain. J. Neurosci. Methods. 2009;184:199–205. doi: 10.1016/j.jneumeth.2009.08.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Novoselov KS, et al. Electric Field Effect in Atomically Thin Carbon Films”. Science. 2004;306(5696):666–669. doi: 10.1126/science.1102896. [DOI] [PubMed] [Google Scholar]
- 8.Narayan, J. & Bhaumik, A. Novel Phase of Carbon, Ferromagnetism, and Conversion into Diamond. Journal of Applied Physics. 118(21), 215303 (2015).
- 9.Hu, M., et al. Compressed Glassy Carbon: An Ultrastrong and Elastic Interpenetrating Graphene Network. Science Advances, E1603213, (2017). [DOI] [PMC free article] [PubMed]
- 10.Ievins. A, Moritz CT. Therapeutic Stimulation for Restoration of Function After Spinal Cord Injury. Physiology (Bethesda) 2017;32:391–8. doi: 10.1152/physiol.00010.2017. [DOI] [PubMed] [Google Scholar]
- 11.Bucher, E. S. et al. A Flexible Software Platform for Fast-Scan Cyclic Voltammetry Data Acquisition and Analysis. Anal Chem., Nov 5 85(21), 10344–53 10.1021/ac402263x (2013). [DOI] [PMC free article] [PubMed]
- 12.Song W, Truong DQ, Bikson M, Martin JH. Trans-Spinal Direct Current Stimulation Immediately Modifies Motor Cortex Sensorimotor Maps. Brain Stimulation: Basic, Translational, and Clinical Research in Neuromodulation. 2017;10(1):e18. doi: 10.1016/j.brs.2016.11.076. [DOI] [Google Scholar]
- 13.Winnie J, Yoshida K, Hofmann UG. In-vivo implant mechanics of flexible, silicon-based ACREO microelectrode arrays in rat cerebral cortex. IEEE Transactions on Biomedical Engineering. 2006;53(5):934–940. doi: 10.1109/TBME.2006.872824. [DOI] [PubMed] [Google Scholar]
- 14.Ansaldo A, Castagnola E, Maggiolini E, Fadiga L, Ricci D. Superior Electrochemical Performance of Carbon Nanotubes Directly Grown on Sharp Microelectrodes. ACS Nano. 2011;5(3):2206–2214. doi: 10.1021/nn103445d. [DOI] [PubMed] [Google Scholar]
- 15.Castagnola E, Ansaldo A, Fadiga L, Ricci D. Chemical vapour deposited carbon nanotube coated microelectrodes for intracortical neural recording. physica status solidi (b) 2010;247(11–12):2703–2707. doi: 10.1002/pssb.201000217. [DOI] [Google Scholar]
- 16.Wang K, Fishman HA, Dai H, Harris JS. Neural Stimulation with a Carbon Nanotube Microelectrode Array. Nano Lett. 2006;6(9):2043–2048. doi: 10.1021/nl061241t. [DOI] [PubMed] [Google Scholar]
- 17.Lu, Y., Lyu, H., Richardson, A. G., Lucas, T. H. & Kuzum, D. Flexible Neural Electrode Array Based-on Porous Graphene for Cortical Microstimulation and Sensing. Scientific Reports, 6:33526, 10.1038/srep33526. [DOI] [PMC free article] [PubMed]
- 18.Cogan SF. Neural stimulation and recording electrodes. Annu. Rev. Biomed. Eng. 2008;10:275–309. doi: 10.1146/annurev.bioeng.10.061807.160518. [DOI] [PubMed] [Google Scholar]
- 19.Cui X, Martin DC. Fuzzy gold electrodes for lowering impedance and improving adhesion with electrodeposited conducting polymer films. Sens. Actuators. 2003;103:384–399. doi: 10.1016/S0924-4247(02)00427-2. [DOI] [Google Scholar]
- 20.19 Macdonald, J. R. Impedance Spectroscopy: Emphasizing solid materials and systems (Wiley, 1987).
- 21.Parker, J. R. et al. Testing of thin-film electrode arrays for cochlear implants of the future, Biomedical Research in the 3rd Millennium. Proceedings of the Inaugural Conference of the Victorian chapter of the IEEE Engineering in Medicine and Biology Society, Caulfield, Victoria (1999)
- 22.Samanta PV, Gonçalves F, Freitas MMA, Pereira MFR, Figueire JL. Surface Activation of a Polymer Based Carbon” Carbon. 2004;42(7):1321–1325. [Google Scholar]
- 23.Goshi N, Vomero N, Dryg I, Kassegne S. Modeling and Characterization of Tissue/Electrode Interface in Capacitive µECoG Glassy Carbon Electrodes. ECS Trans. 2016;72(1):83–90. doi: 10.1149/07201.0083ecst. [DOI] [Google Scholar]
- 24.Reddy VM, et al. Determination of dopamine in presence of ascorbic acid and uric acid using poly (Spands Reagent) modified carbon paste electrode. Mater Sci Eng C Mater Biol Appl. 2015;57:378–86. doi: 10.1016/j.msec.2015.08.005. [DOI] [PubMed] [Google Scholar]
- 25.Hirabayashi M, Mehta B, Vahidi NW, Khosla A, Kassegne S. Functionalization and characterization of pyrolyzed polymer based carbon microstructures for bionanoelectronics platforms. Journal of Micromechanics and Microengineering. 2013;23(11):115001. doi: 10.1088/0960-1317/23/11/115001. [DOI] [Google Scholar]
- 26.Castagnola E, Biso M, Ricci D. Improvement of polypyrrole and carbon nanotube co-deposition techniques for high charge transfer electrodes. Phys. Status Solidi (B). 2009;246:2469–2472. doi: 10.1002/pssb.200982283. [DOI] [Google Scholar]
- 27.Castagnola V, Bayon C, Descamps E, Bergaud C. Morphology and conductivity of PEDOT layers produced by different electrochemical routes. Synth. Met. 2014;189:7–16. doi: 10.1016/j.synthmet.2013.12.013. [DOI] [Google Scholar]
- 28.Castagnola, E. et al. Ultra-flexible and brain-conformable micro-electrocorticography device with low impedance PEDOT-carbon nanotube coated microelectrodes, Neural Engineering (NER), 2013 6th International IEEE/EMBS Conference on, 927–930 (2013).
- 29.Venkatraman S, et al. In Vitro and In Vivo Evaluation of PEDOT Microelectrodes for Neural Stimulation and Recording. IEEE Trans Neural Syst Rehabil Eng. 2011;19(3):307–16. doi: 10.1109/TNSRE.2011.2109399. [DOI] [PubMed] [Google Scholar]
- 30.Beebe X, Rose TL. Charge injection limits of activated iridium oxide electrodes with 0.2 ms pulses in bicarbonate buffered saline. IEEE Trans. Biomed. Eng. 1988;35:494–95. doi: 10.1109/10.2122. [DOI] [PubMed] [Google Scholar]
- 31.Kassegne SK, et al. Numerical Modeling of Transport and Accumulation of DNA on Electronically Active Biochips. Journal of Sensors and Actuators B: Chemical. 2003;94:81–98. doi: 10.1016/S0925-4005(03)00322-8. [DOI] [Google Scholar]
- 32.Johnson LA, et al. Direct electrical stimulation of somatosensory cortex in humans using electrocorticography electrodes: a qualitative and quantitative report. J. Neural Eng. 2013;10:036021. doi: 10.1088/1741-2560/10/3/036021. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Ordonez JS, Boehler C, Schuettler M, Stieglitz T. Long-term adhesion studies of polyimide to inorganic and metallic layers. MRS Online Proceeding. 2012;1466:7–13. [Google Scholar]
- 34.Diaz-Botia C, et al. A silicon carbide electrode technology for the central and the peripheral nervous system. J. Neural Eng. 2017;14(5):056006. doi: 10.1088/1741-2552/aa7698. [DOI] [PubMed] [Google Scholar]
- 35.Arreaga-Salas DE, et al. Integration of High-Charge-Injection-Capacity Electrodes onto Polymer Softening Neural Interfaces. ACS Appl. Mater. Interfaces. 2015;7(48):26614–26623. doi: 10.1021/acsami.5b08139. [DOI] [PubMed] [Google Scholar]
- 36.Vomero, M. et al. Incorporation of Silicon Carbide and Diamond-Like Carbon as Adhesion Promoters Improves In Vitro and In Vivo Stability of Thin-Film Glassy Carbon ECoG Arrays, Advanced Biosystems, accepted October 2017. 10.1002/adbi.201700081.
- 37.Weiland JD, Anderson DJ, Humayun MS. In vitro electrical properties for iridium oxide versus titanium nitride stimulating electrodes. Biomedical Engineering, IEEE Transactions on. 2002;49:1574–1579. doi: 10.1109/TBME.2002.805487. [DOI] [PubMed] [Google Scholar]
- 38.Cogan SF, Guzelian AA, Agnew WF, Yuen TG, McCreery DB. Over-pulsing degrades activated iridium oxide films used for intracortical neural stimulation. Journal of Neuroscience Methods. 2004;137:141–150. doi: 10.1016/j.jneumeth.2004.02.019. [DOI] [PubMed] [Google Scholar]
- 39.Cui XT, Zhou DD. Poly (3,4-ethylenedioxythiophene) for chronic neural stimulation. IEEE Trans Neural Syst Rehabil Eng. 2007;15(4):502–508. doi: 10.1109/TNSRE.2007.909811. [DOI] [PubMed] [Google Scholar]
- 40.Wilks SJ, Richardson-Burns SM, Hendricks JL, Martin DC, Otto KJ. Poly (3,4-ethylenedioxythiophene) as a microneural interface material for electrostimulation. Frontiers in Neuroengineering. 2009;2:7. doi: 10.3389/neuro.16.007.2009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Rose T, Kelliher E, Robblee L. Assessment of capacitor electrodes for intracortical neural stimulation. Journal of Neuroscience Methods. 1985;12:181–193. doi: 10.1016/0165-0270(85)90001-9. [DOI] [PubMed] [Google Scholar]
- 42.Jan E, et al. Layered carbon nanotube-polyelectrolyte electrodes outperform traditional neural interface materials. Nano Letters. 2009;9:4012–4018. doi: 10.1021/nl902187z. [DOI] [PubMed] [Google Scholar]
- 43.Castagnola E, et al. PEDOT-CNT-Coated Low-Impedance, Ultra-Flexible, and Brain-conformable Micro-ECoG Arrays. IEEE Transactions on Neural Systems and Rehabilitation Engineering. 2015;23:342–350. doi: 10.1109/TNSRE.2014.2342880. [DOI] [PubMed] [Google Scholar]
- 44.Bullara L, McCreery D, Yuen T, Agnew W. A microelectrode for delivery of defined charge densities. Journal of Neuroscience Methods. 1983;9:15–21. doi: 10.1016/0165-0270(83)90104-8. [DOI] [PubMed] [Google Scholar]
- 45.Prasad A, et al. Abiotic-biotic characterization of Pt/Ir microelectrode arrays in chronic implants, Frontiers in Neuroengineering. 2014;4(7):2. doi: 10.3389/fneng.2014.00002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Kozai TDY, et al. Chronic In Vivo Evaluation of PEDOT/CNT for Stable Neural Recordings. IEEE Trans Biomed Eng. 2016;63(1):111–119. doi: 10.1109/TBME.2015.2445713. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Huynh, N. U. et al. Effect of Thin Film Platinum Microelectrode Granularity on Corrosion under Electrical Stimulation, ECS Meeting Abstracts, 1064–106.
- 48.Patrick E, Orazem ME, Sanchez JC, Nishida T. Corrosion of tungsten microelectrodes used in neural recording applications. Journal of Neuroscience Methods. 2011;198(2):158–171. doi: 10.1016/j.jneumeth.2011.03.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Jackowska K, Krysinsky P. New trends in the electrochemical sensing of dopamine. Anal Bioanal Chem. 2013;405(11):3753–3771. doi: 10.1007/s00216-012-6578-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Yang C, et al. Carbon Nanotubes Grown on Metal Microelectrodes for the Detection of Dopamine. Anal. Chem. 2016;88:645–652. doi: 10.1021/acs.analchem.5b01257. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Xiao N, Venton BJ. Rapid, Sensitive Detection of Neurotransmitters at Microelectrodes Modified with Self-assembled SWCNT Forests. Anal Chem. 2012;84(18):7816–7822. doi: 10.1021/ac301445w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Taylor IM, et al. Enhanced Dopamine Detection Sensitivity by PEDOT/Graphene Oxide Coating on in vivo Carbon Fiber Electrodes. Biosens Bioelectron. 2017;89(Pt 1):400–410. doi: 10.1016/j.bios.2016.05.084. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Yang C, et al. Laser Treated Carbon Nanotube Yarn Microelectrodes for Rapid and Sensitive Detection of Dopamine in Vivo, ACS Sens. 2016;1:508–515. doi: 10.1021/acssensors.6b00021. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Yi W, et al. 3D Carbon Nanofiber Microelectrode Arrays Fabricated by Plasma-Assisted Pyrolysis to Enhance Sensitivity and Stability of Real-time Dopamine Detection. Biomed Microdevices. 2016;18:112. doi: 10.1007/s10544-016-0136-1. [DOI] [PubMed] [Google Scholar]
- 55.Zachek MK, Takmakov P, Moody B, Wightman RM, McCarty GS. Simultaneous Decoupled Detection of Dopamine and Oxygen using Pyrolyzed Carbon Microarrays and FSCV. Anal Chem. 2009;81(15):6258–6265. doi: 10.1021/ac900790m. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Bath BD, et al. Sub-second Adsorption and Desorption of Dopamine at Carbon-Fiber Microelectrodes. Anal. Chem. 2000;72(24):5994–6002. doi: 10.1021/ac000849y. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.