Skip to main content
. 2018 Jan 24;28(6):2464–2473. doi: 10.1007/s00330-017-5113-4

Table 3.

Dependency on tube load for different reconstruction algorithms. Visual grading regression (VGR) coefficients for log (mAs) in pairwise comparisons of two quality reference (Qref) mAs values

Criterion Reconstruction algorithm
FBP ADMIRE 3 ADMIRE 5
42 mAs vs. 98 mAs 98 mAs vs. 140 mAs 42 mAs vs. 98 mAs 98 mAs vs. 140 mAs 42 mAs vs. 98 mAs 98 mAs vs. 140 mAs
1. Liver parenchyma 3.36** 2.20*** –0.28° 1.49* –0.37°
2. Pancreatic contours 1.89** 2.41*** –0.89* 1.37*** –2.62***
3. Kidneys and proximal ureters 7.25* 2.39* 2.29*** –0.92* 1.46*** –2.20***
4. Lymph nodes < 15 mm in diameter 6.50* 2.09** 1.83*** –1.93*** 1.14*** –2.20*
5. Image noise 5.49** 4.70** 3.07** 4.79* 1.07°
6. Overall image quality 3.68* 0.03° 3.74** –1.27**

* p < 0.05, ** p < 0.01, *** p < 0.001, ° not significant

– Convergence was not achieved with the statistical software available