Abstract
Objective
Treatment with intrathecal baclofen (ITB) is commonly used in patients with severe spasticity. However, complications may occur after implantation of the ITB‐device, albeit mainly procedure‐ and device‐related problems. The aim of the study was to assess surgical‐ as well as catheter‐ and pump‐related complications and define their risk factors.
Methods
We retrospectively evaluated all patients with an implanted ITB‐device who were treated at the Department of Neurology, Hochzirl Hospital, Zirl, Austria, between 2006 and 2016.
Results
Twenty‐nine of 116 (25%) patients experienced 32 complications: 5 procedure‐ and 27 device‐related (4 pump‐ and 23 catheter‐associated) problems occurred. Risk factors for sustaining any complication were a spinal localization of lesion (odds ratio [OR] OR 2.71, p = .021), other catheter types than an Ascenda® catheter (OR 3.87, p = .041), a lower modified Rankin Scale (median 4 vs. 5; OR 2.86, p = .015) and a higher Barthel Index (median 53 vs. 0; OR 2.84, p = .006). The median time from the last ITB‐related surgery to the first complication was 18 (IQR 1‐57) months. Overall, 47% complications occurred within the first year after any surgical procedure regarding the ITB‐device, thereof 25% within the first month.
Conclusions
Procedure‐ and device‐related complications are frequent after implantation of an ITB‐device with catheter‐associated complications as the most frequently encountered problems. Patients with a spinal origin of spasticity, a lower modified Rankin Scale and a higher Barthel Index have a higher risk to sustain a complication.
Keywords: autonomic nervous system diseases, baclofen, brain injuries, sympathetic nervous system, traumatic
1. INTRODUCTION
Intrathecal baclofen (ITB) has successfully been used in patients with severe spasticity of different etiologies for more than 30 years (Albright, Barron, Fasick, Polinko, & Janosky, 1993; Albright, Cervi, & Singletary, 1991; Armstrong et al., 1997; Becker, Alberti, & Bauer, 1997; Becker, Sure, Petermeyer, & Bertalanffy, 1999; Coffey et al., 1993; Dario, Di Stefano, Grossi, Casagrande, & Bono, 2002; Meythaler, DeVivo, & Hadley, 1996; Meythaler, Guin‐Renfroe, Grabb, & Hadley, 1999; Meythaler, McCary, & Hadley, 1997; Penn & Kroin, 1985; Penn et al., 1989; Rawicki, 1999; Rifici et al., 1994). Treatment of severe spinal as well as supraspinal spasticity with oral antispastic medication often results in an insufficient response and is limited by intolerable side effects. The therapeutic utility of ITB in the management of severe spasticity has been repeatedly demonstrated, but nevertheless, complications may occur after the ITB‐pump has been implanted and therefore decrease therapeutic effects. The classification often used in the literature comprises the differentiation between procedure‐related, device‐related and drug‐related complications (Borrini et al., 2014; Turner, Sears, & Loeser, 2007). A review made in 2010 only focused on procedure‐ and device‐related complications of ITB‐administration (Stetkarova, Yablon, Kofler, & Stokic, 2010): 558 complications were reported after 1,362 pump implantations, of which 27% were related to surgical procedures, 7% to pump problems, and 66% to catheter malfunctions.
In a prospective, observational cohort study in 158 adults treated with ITB therapy and followed up for 1 year, 18% experienced 38 adverse events (Borrini et al., 2014). Motta and coworkers analyzed complications and risk factors in 200 consecutive children and adolescents after intrathecal baclofen pump implantation (Motta, Buonaguro, & Stignani, 2007). A statistically significant correlation between the occurrence of a complication and patients with an Ashworth Scale higher than 3 and an age of 10 years or younger was found. In the largest series in 430 consecutive children treated with ITB in a 14‐year period, 137 complications occurred in 25% (Motta & Antonello, 2014).
The purpose of this study is to give a detailed account of the most frequently encountered complications, catheter‐ and pump‐associated problems and define their risk factors.
2. MATERIALS AND METHODS
We retrospectively evaluated all patients with an intrathecal ITB‐device, who were treated at our Department of Neurology, between 01.01.2006 and 31.12.2016. Patients of any age and receiving care bothin inpatient as well as in outpatient hospital setting were included.
ITB‐pump‐related surgical interventions were classified as “primary implantation procedure” (pump implantation for the first time) and “replacement procedure” (pump surgery: replacement due to end of battery life 72 months after implantation at the earliest or revision because of a complication; catheter‐only‐revision after a complication; exchange of the entire system) according to earlier published studies (Awaad et al., 2012; Borrini et al., 2014). The retrospective analysis was done by reviewing patient's medical history and the data provided in our ITB‐dedicated register, which was implemented in 01.01.2006. Exclusion criteria were insufficient patient data. The majority of patients were admitted for ITB‐evaluation, presurgical and postsurgical care as well as follow‐up after implantation of the ITB‐pump. The remaining patients were referred because of an occurring complication for diagnostic work‐up and treatment but followed up at their local district hospital. According to earlier published studies, complications were divided into procedure‐related and device‐related complications, the latter into catheter‐ and pump‐associated complications (Guillaume, Van Havenbergh, Vloeberghs, Vidal, & Roeste, 2005; Kolaski & Logan, 2007; Rawlins, 2004). Procedure‐associated problems were defined as related to the surgical intervention occurring within the first 2 months after a surgical intervention and thus most likely being associated with the surgical intervention, such as scar complications, subcutaneous seroma, infections, or cerebrospinal fluid (CSF) leakages. Device‐associated problems include complications related to the catheter or the pump and were classified as “early” when they occurred within the first 2 months or “late” when occurring later than 2 months after the surgical intervention.
We collected clinical data in all included patients: demographic information, spinal versus supraspinal localization of lesion responsible for the spasticity and data of any previous ITB‐device‐related surgery. Moreover, the type of the implanted catheter (Ascenda® catheter/Model 8781 vs. other catheter types) and the following parameters at steady‐state of ITB were assessed: Ashworth Scale, modified Rankin Scale, Barthel Index, the median ITB‐dosage, median ITB‐concentration, flow rate of ITB, mode of application, flow rate of ITB, and last follow‐up. In patients admitted more than once, the follow‐up was defined as period in months between the first and last consultation at our department. In patients with complications, we analyzed the number and type of complications, as well as the type of intervention. Patients with a suspected complication underwent a comprehensive diagnostic work‐up including the evaluation of the clinical response to ITB, laboratory test and imaging (depending on the clinical situation conventional x‐ray, computer tomography without or with contrasting agent, magnet resonance imaging or fluoroscopy). If the aetiology of the complication still remained unclear, a surgical exploration was done. In patients with a complication, we calculated the duration of the hospital stay from admission to the day of a consecutive intervention (revision surgery, surgical exploration, or nonsurgical intervention). Moreover, the time from the initial ITB‐device implantation and from the last ITB‐related surgery respectively to any occurring complication during the observational period was calculated in all patients with complications. In addition, risk factors for sustaining a complication were analyzed.
2.1. Statistical analysis
Data were summarized in cross tables, and medians and interquartile ranges were calculated for ordinal variables, such as modified Rankin Scale, Barthel Index, or Asworth Scale. Complication and incident rates for the occurrence of complications were calculated and odds ratios comparing different subgroups regarding complication frequency were given. Since there were three patients with two complications, statistical analysis was conducted with cases including the second complication. Odds ratios for ordinal risk factors were obtained by dichotomizing the respective risk factor using the overall median value as the cut‐off threshold. The influence of risk factors on the occurrence of complications was tested by chi‐squared test for nominal and by Mann–Whitney U test for ordinal risk factors. A significance level of α = .05 (2‐tailed) was applied. Statistical analyses were performed using SPSS, version 22.0 (IBM Corp., Armonk, NY, USA).
According to our country's law on retrospective research, this study did not require the approval of the ethics committee, but was performed based on ethical standards of the “World Medical Association Helsinki Declaration” (https://www.wma.net/policies-post/wma-declaration-of-helsinki-ethical-principles-for-medical-research-involving-human-subjects/).
3. RESULTS
One hundred and thirty‐six patients with an ITB‐pump were treated within the given 11‐year period at our neurorehabilitation department. One hundred and sixteen patients (77 men and 39 women, mean age at implantation 39 ± 16.7 years, range 2‐77 years) entered the final analysis. The location of lesion responsible for the spasticity was supraspinal in 73/116 (60%) patients with traumatic etiology in 34/116, hypoxic in 16/116, and 23/116 had other etiologies (seven infantile cerebral palsy, six intracerebral hemorrhage, four ischemic lesions, three subarachnoidal hemorrhage, one stiff person syndrome, one extra‐/pontine myelinolysis, one undetermined neurodegenerative disease). Forty‐three of 116 (40%) patients had a spinal lesion as origin of spasticity (18 traumatic, 14 multiple sclerosis, four sporadic spastic paralysis, three hereditary spastic paralysis, one disc herniation, one ischemic lesion, one myelopathy due to extradural granuloma, one undetermined spastic paraplegia). Patients’ characteristics and further clinical information are displayed in Table 1.
Table 1.
Patient characteristics | |
Age at implantation (mean ± standard deviation; years) | 39 ± 16.7 |
Sex (women/men; n) | 39/77 |
Pathologic disorders | |
Supraspinal [n/%] | |
Traumatic brain injury | 34/29% |
Cerebral hypoxia | 16/14% |
Cerebral palsy | 7/6% |
Intracerebral hemorrhage | 6/5% |
Ischemic stroke | 4/3% |
Subarachnoidal hemorrhage | 3/3% |
Stiff person syndrome | 1/1% |
Extra‐/pontine myelinolysis | 1/1% |
Unknown neurodegenerative disease | 1/1% |
Spinal [n/%] | |
Traumatic spinal injury | 18/15% |
Multiple sclerosis | 14/12% |
Spastic paralysis | 7/6% |
Disc herniatio | 1/1% |
Ischemic lesion | 1/1% |
Myelopathy due to extradural granuloma | 1/1% |
Spastic paresis of unclear etiology | 1/1% |
Investigated parametersa | |
Median modified Rankin Scale (IQR) | 5 (4‐5) |
Median Barthel Index (IQR) | 0 (0‐54) |
Median ITB‐dosage (IQR) [μg/day] | 200 (129‐320) |
Median ITB‐concentration (IQR) [μg/ml] | 2,000 (500‐2,000) |
Median flow rate (IQR) [μl/hr] | 6.3 (3.5‐10.4) |
Median Ashworth Scale (IQR) | 4 (3‐4) |
Catheter type [Ascenda®, Model 8,781/other types; n/%] | 39/77 |
IQR, interquartile range; ITB, intrathecal baclofen.
The medians of the investigated parameter were calculated on basis of the patient population.
One hundred and eight of 116 primary implantation procedures as well as surgical management of all complications were performed at the Department of Neurosurgery, 8/116 ITB‐pumps were implanted in another neurosurgical department. Overall, 143 surgical procedures were performed period within the 11‐year observational period: 82 implantations of the ITB‐pump for the first time (the remaining 34 implantations were done before 01.01.2006), 35 replacement‐surgeries due to end of battery life (one time in 29 patients, two times in three patients) and 26 revision procedures due to a complication (two times in three patients each). Seventy‐seven intrathecal catheter types other than an Ascenda® (Model 8781) catheter were implanted. The median Ashworth Scale of all included patients at steady‐state of ITB was 4 (interquartile range [IQR] 3‐4), the median modified Ranking Scale 5 (IQR 4‐5) and the median Barthel Index 0 (IQR 0‐54). The median ITB‐dosage was 200 μg/d (IQR 129‐320) at steady‐state of ITB, the median ITB‐concentration 2000 μg/ml (IQR 500‐2000; 82/116 2000 μg/ml, 30/116 500 μg/ml, 2/116 1000 μg/ml, 1/116 1250 μg/ml, 1/116 250 μg/ml) and the median flow rate 6.3 μl/hr (IQR 3.5‐10.4) (see also Table 1). The mode of application after reaching steady‐state of ITB was continuous in 95% and flex mode in 5% patients. The median follow‐up was 42 months (IQR 17‐91, mean 56 months), comprising 506 pump years. Seven patients died during follow‐up (one suicide, three severe pneumonia, three sepsis). There was no permanent morbidity or death related to the ITB‐pump‐complication.
3.1. Complications
Thirty‐two complications occurred in 29/116 patients during the 11‐year study period. Three patients experienced two complications each. Five of 32 (16%) complications were related to surgical procedures and 27/32 (84%) were device‐related (23 catheter‐associated problems, four pump‐associated problems). Procedure‐related complications contained two pump site infections and three CSF leakages. Catheter‐related problems were: nine dislocations (in eight patients), three CSF leakages, five disconnections, three kinks, one break, and two unknown catheter dysfunctions. Pump‐related problems were caused by hypermobility (1/32) and by pump site infections (3/32 in three patients). Six of 27 device‐associated problems were early complications (three CSF leakages, three pump site infections), the remaining 21/27 were late device‐related complications.
The diagnostic work‐up resulted in 24/29 patients with complications in 27 revision procedures: 11 exchanges of the entire device in 11 patients, 14 revisions of the catheter in 13 patients (nine entire catheter, four distal segment, one swap of proximal segment) and two exchanges of the pump in two patients. In three patients a nonsurgical intervention was performed (application of blood patches in two patients and abdominal compression in one patient). In two patients, a surgical exploration revealed no obvious reason for the dysfunction of the entire system; thereafter, in one patient the device functioned normally again (thus catheter kinking suspected as most likely etiology), in the other patient the ITB‐pump was explanted after an interdisciplinary decision due to functional worsening of a hemispastic syndrome. In patients with a complication, the median duration of the hospital stay from admission to the consecutive interventional procedure to resolve the complication was 5 (IQR 2‐13.75) days. Further details of patients with a complication are given in Table 2.
Table 2.
No | Pathology | Complication | Time to eventb [days] | Imaging | Procedure | Median modified Rankin Scale | Median barthel index | Median flow rate [μl/hr] | Median ashworth scale | Catheter |
---|---|---|---|---|---|---|---|---|---|---|
1 | SP | Catheter dislocation of distal segment | 64 | XR: dislocation | Exchange of entire catheter | 2 | 95 | 9.2 | 4 | 8731SC |
2 | TBI | Catheter dislocation of distal segment | 914 | CT: dislocation | Exchange of entire catheter | 5 | 0 | 5.8 | 4 | 8731SC |
3 | TSI | CSF leakage (early) | 24 | No imaging | Patching | 3 | 80 | 25.4 | 4 | 8731SC |
4 | Hypoxia | CSF leakage (late) | 1,822 | No imaging | Exchange of entire system | 4 | 60 | 18.6 | 4 | 8709/8596 |
5a | Disc herniation | Catheter disconnection | 737d | No imaging | Exchange of entire catheter | 4 | 65 | 10.8 | 0 | 8709/8596 |
6a | Disc herniation | Pump site infection | 884d | No imaging | Exchange of pump | 4 | 65 | 5.1 | 0 | 8709/8596 |
7 | ICH | Pump site infection | 168 | No imaging | Explantation of pump | 4 | 45 | 6.7 | 4 | 8731SC |
8 | TBI | Catheter kinking | 19c | XR+CA: no signs | No revision | 4 | 0 | 4.6 | 4 | 8731SC |
9 | Spinal ischemia | Unknown catheter dysfunction | 2,471 | Fluoroscopy: No signs | Exchange of entire system | 4 | 30 | 7.1 | 4 | 8731SC |
10 | TBI | Catheter kinking | 115 | XR+CA: no signs | Exchange of distal segment | 5 | 0 | 22.9 | 5 | 8711 |
11a | TSI | Catheter dislocation of distal segment | 32 | No imaging | Exchange of entire catheter | 4 | 75 | 19.2 | 3 | 8731SC |
12a | TSI | Catheter dislocation of distal segment | 28d | XR: no signs | Exchange of entire catheter | 4 | 75 | 19.2 | 3 | 8731SC |
13 | MS | Catheter kinking | 364 | XR & CT+CA: no signs | Exchange of entire catheter | 5 | 20 | 2.7 | 4 | 8781 |
14 | TSI | Catheter dislocation of distal segment | 657 |
XR: catheter at L3 CT+CA: no signs |
Exchange of distal segment | 4 | 75 | 2.7 | 1 | 8731SC |
15 | CP | Pump site infection | 26 | No imaging | Exchange of entire system | 5 | 0 | 4.2 | 4 | 8781 |
16 | Hypoxia | Pump site infection | 368 | No imaging | Exchange of entire system | 5 | 0 | 8.3 | 5 | 8731SC |
17 | SAH | Catheter dislocation of distal segment | 25 |
CT+CA: dislocation Fluoroscopy: dislocation |
Exchange of distal segment | 5 | 0 | 3.1 | 4 | 8711 |
18 | ICH | Catheter disconnection | 2,470d | XR: disconnection | Exchange of pump and proximal segment | 4 | 40 | 2.3 | 4 | 8709/8596 |
19 | TBI | Catheter dislocation of distal segment | 838d | Fluoroscopy: no signs | Exchange of entire catheter | 5 | 0 | 2.3 | 5 | 8711 |
20 | SP | Unknown catheter dysfunction | 1,174 | MRI: no signs | Exchange of distal segment | 1 | 100 | 22.5 | 5 | 8781 |
21 | Hypoxia | Pump site infection | 23 | No imaging | Exchange of entire system | 5 | 0 | 39.2 | 5 | 8781 |
22 | MS | Catheter disconnection of proximal segment | 33 | CT: no signs | Exchange of proximal segment | 5 | 5 | 7.5 | 5 | 8781 |
23 | Hypoxia | CSF leakage (early) | 16 | No imaging | Extracorporal compression | 5 | 0 | 6.3 | 4 | 8731SC |
24 | TSI | Catheter disconnection | 1,864d | XR & CT: no signs | Exchange of entire catheter | 4 | 75 | 11.9 | 0 | 8711 |
25 | SP | Catheter break | 1,570d | CT+CA: no intradural contrast | Exchange of entire system | 2 | 85 | 14.6 | 4 | 8731SC |
26 | SP | CSF leakage (early) | 44 | MRI: leakage | Patching | 1 | 100 | 2.7 | 3 | 8781 |
27 | TBI | Catheter dislocation of distal segment | 2,058 | CT+CA: no signs | Exchange of entire system | 5 | 0 | 7.2 | 5 | 8731SC |
28 | SP | Catheter disconnection | 2,179d |
XR: disconnection CT: loop |
Exchange of entire system | 2 | 100 | 11.0 | 3 | 8731SC |
29a | TSI | CSF leakage (late) | 2,200d |
XR + CA: no signs CT + CA: leakage |
Exchange of entire system | 4 | 60 | 5.2 | 0 | 8731SC |
30a | TSI | Pump hypermobility | 1,683c | No imaging | Pump exchange | 4 | 67.5 | 10 | 0 | 8731SC |
31 | TSI | Catheter dislocation of distal segment | 12d | No imaging | Exchange of entire catheter | 4 | 80 | 11.2 | 0 | 8731SC |
32 | SAH | CSF leakage (late) | 2,270 | XR: no signs | Exchange of pump and proximal segment | 4 | 10 | 8.3 | 3 | 8731SC |
CA, contrasting agent; CP, cerebral palsy; CSF, cerebrospinal fluid; CT, computer tomography; MRI, magnet resonance imaging; MS, multiple sclerosis; No, number of complication; SAH, subarachnoidal hemorrhage; SP, spastic paralysis; TBI, traumatic brain injury; TSI, traumatic spinal injury; XR, conventional x‐ray.
2 complications in 3 patients each.
Time to event was calculated from the last surgical procedure.
First implantation or revision surgery.
Scheduled device swap due to end of battery life to the date of a consecutive intervention (surgical or non‐surgical interventions).
The complication rate after implantation of an ITB‐pump was 0.063 per pump year for any type of complication in the 11‐year follow‐up of our patient group, for surgical‐related complications 0.01 per pump year and for device‐related problems 0.053 per pump year respectively. Per implantation, the complication rate was 0.24. Twenty of 32 complications occurred after the first implantation of the device, and 12/32 complications after scheduled replacement surgeries due to end of battery life or replacement surgeries due to revision. Here, statistical analysis revealed no significant increase in complications after the initial implantation of the device compared to replacement surgeries. The median time from the last ITB‐related surgery to the first complication was 18 (IQR 1‐57) months. Overall, 15/32 (47%) complications occurred within the first 12 months after the last surgical intervention regarding the ITB system, thereof 8/32 (25%) complications in the first month (5 after first implantation, 3 after replacement) (Figures 1 and 2).
3.2. Risk factors for complications
Analysis revealed a significant association between occurrence of a complication and a spinal localization of lesion responsible for the spasticity (OR 2.71, p = .021), whereas sex (p = .663, age (p = .304) were not significantly related to a complication. Furthermore, there was a significant difference regarding the independence in activities of daily life in patients with complications compared to patients without complications (Barthel Index 53 [IQR 0‐75] vs. 0 [IQR 35‐0]; OR 2.84, p = .006). Patients with a complication also differed significantly in the modified Rankin Scale from patients without a complication (4 [IQR 4‐5] vs. 5 [IQR 4‐5]; OR 2.86, p = .015). No significant differences in both patient groups were found regarding spasticity (Ashworth Scale 4 vs. 3, p = .381) and flow rate (5.6 μl/hr vs. 7.9, p = .065).
A subgroup analysis on the 21 patients with 22 catheter‐related complications revealed that other catheter types than an Ascenda® catheter (introduced in 2010, Model 8781) were also identified as a risk factor (19/73, 26% vs. 3/36, 8%; OR 3.87, p = .041).
4. DISCUSSION
This study provides a systematic assessment of procedure‐ and device‐related complications after implantation of an ITB‐pump in a large patient cohort. Twenty‐five per cent of the patients had complications, most frequently catheter‐associated problems in our series, followed by procedure‐ and pump‐related complications. Risk factors were a spinal lesion, older catheter types, a lower modified Rankin Scale and a higher Barthel Index. About half of the complications occurred within the first 12 months after any ITB‐related surgery, half of these occurring within the first month.
Comparison of our results with previously published papers is limited due to the variety of different methodologies, study‐populations, included types of adverse events and duration of follow‐up. The review of 32 full‐length manuscripts and 10 case reports published by Stetkarova and co‐workers in 2010 also focused exclusively on procedure‐, pump‐, and catheter‐related complications of ITB‐administration (Stetkarova et al., 2010). Complication rates varied widely among the studies with a mean complication rate of 0.41 per implantation but ranging from 0‐2.24 per implantation across the reviewed studies. The complication rate after implantation of an ITB‐pump was 0.24 per implantation in our study cohort, thus lying within the reported range. The authors found higher complication rates in centers with a longer mean follow‐up than 18 months (0.56 ± 0.56 vs. 0.23 ± 0.19). With regard to our mean follow‐up of 56 months, our complication rate lies at the lower part of the range.
Based on the results of our study, we found a nearly threefold increased risk of a complication in patients with a spinal localization of lesion responsible for the spasticity versus supraspinal localization. This is in contrast with Borrini and coworkers who reported no significant correlation between localization of lesion and occurrence of adverse events (Borrini et al., 2014). The reason for this discordant finding is not clear, possibly methodological differences might play a role as the study period was only 1 year and median follow‐up duration was substantially shorter (10 months vs. 42 months in our cohort). In our study group, a higher Barthel Index regarding the independence in activities of daily life and a lower modified Rankin Scale related to the ambulatory status were both associated with a nearly threefold increased risk for sustaining a complication. This is also in contrast with Borrini and coworkers who did not find a significant correlation with the ambulatory status. The results of our study might be supported by findings of a Russian working group analysing complications in 12/52 (23%) patients after ITB‐device implantation (Paskhin, Dekopov, Tomsky, Isagulyan, & Salova, 2017): 5/12 complications were caused by catheter migration and the authors identified patients with severe dystonia of the trunk muscles to have an increased risk of spinal catheter migrations. Accordingly, the degree of independence and mobility represented with the Barthel Index and the modified Rankin Scale might be correlated with the occurrence of catheter problems based on mechanical stress.
In our cohort, no significant association was found between occurrence of an adverse event and gender, which is in line with a previous report (Borrini et al., 2014). Also age was no risk factor for a complication in our study which is in contrast with a previously published paper (Motta et al., 2007): in the study of Motta and coworkers, complications were statistically more likely in patients with an age of 10 years and younger. This might be related to the by far lower mean age of 13.7 ± 5.7 years of the 200 children and adolescents in their study compared to ours (mean age 39 ± 16.7 years). An Ashworth Scale higher than 3 was a further risk factor in this study whereas we did not find a significant difference of the Ashworth Scale in patients with and without complications. The divergent results might be explained by the different patient population, as about 90% of included patients in the study of Motta and coworkers were affected by cerebral palsy with thus clinically dominating dystonia, whereas we included a more mixed patient population with a traumatic etiology in more than half of the patients.
The impact of dose and concentration of intrathecally administered drugs on catheter tip granuloma is still discussed controversially. A direct correlation with high concentrations or high dose opiate drugs was suggested (Coffey & Burchiel, 2002; Yaksh et al., 2002), but also an association between intrathecal granulomas and low doses and concentrations of hydromorphone (Veizi et al., 2016) reported. We analyzed a possible association between the flow rate of ITB and possibly mechanically driven complications and found no relation to occurring complications.
Motta and coworkers compared complications before and after the introduction of the Ascenda® catheter for the first time in paediatric patients recently (Motta & Antonello, 2016): 120/416 (29%) children with silicone catheters sustained major complications (infections, leakages, catheter‐related problems), but only 1/92 children with an Ascenda® catheter (leakage). We could confirm these results in our study as an implanted Ascenda® catheter was associated with an approximately fourfold decreased risk for catheter‐related complications, most likely the consequence of technological improvements of the Ascenda® catheter.
In our cohort, half of the complications occurred within the first 12 months after any ITB‐related surgery, after primary implantation procedures as well as replacement procedures due to end of battery life or revision likewise, within the first month a quarter of the complications. Also Motta and coworkers did not observe a different event rate after initial device‐implantations compared to complications after replacement surgeries (Motta & Antonello, 2014). Thus, the critical period is the first year and especially the first month after any ITB‐related surgery.
4.1. Study limitations
Limitations of our study are: First of all, the retrospective assessment of the data has to be taken into account. Furthermore, the investigated patients of our rehabilitation center during the study period comprise a mixed patient population regarding the different etiologies of spasticity. Moreover, the follow‐up periods varied widely among the patients as all patients with an intrathecal ITB‐device and any treatment at our department (once only admissions as well as regular admissions) during the observational period were included.
Care of patients with an ITB‐device requires a highly specialized, interdisciplinary operating, experienced staff of neurologists, neurosurgeons, nurses, and physiotherapists. Knowledge of adverse events and their risk factors is necessary for patient counseling about risks and benefits before and adequate care after the implantation of the ITB‐device. A spinal localization of lesion, other catheter types than an Ascenda® catheter, a lower modified Rankin Scale and a higher Barthel Index are most predictive for sustaining a complication. Based on the results of our study, the following measures might be taken to avoid as well as identify complications as early as possible: 1. Appropriate care of high‐risk patients i.e. (i) patients with a spinal localization of lesion combined with a higher Bartel Index, possibly related to the ability to participate in the extended activities of daily life and (ii) patients with less impairment of the ambulatory status and therefore, a higher grade of trunk mobility 2. Use of newer catheter types which might be helpful to reduce complications according to improvements of the catheter material. 3. The temporal distribution of complications might be helpful when scheduling postsurgical care and follow‐up after the first implantation of the ITB‐device as well as any replacement surgery.
Further prospective multicenter studies are necessary to evaluate risks and safety issues of treatment with an ITB‐device in order to improve the risk management of complications related to the ITB‐device.
CONFLICT OF INTEREST
None declared.
ACKNOWLEDGMENTS
There are no acknowledgments.
Pucks‐Faes E, Hitzenberger G, Matzak H, et al. Eleven years’ experience with Intrathecal Baclofen – Complications, risk factors. Brain Behav. 2018;8:e00965 https://doi.org/10.1002/brb3.965
REFERENCES
- Albright, A. L. , Barron, W. B. , Fasick, M. P. , Polinko, P. , & Janosky, J. (1993). Continuous intrathecal baclofen infusion for spasticity of cerebral origin. JAMA, 270, 2475–2477. https://doi.org/10.1001/jama.1993.03510200081036 [PubMed] [Google Scholar]
- Albright, A. L. , Cervi, A. , & Singletary, J. (1991). Intrathecal baclofen for spasticity in cerebral palsy. JAMA, 265, 1418–1422. https://doi.org/10.1001/jama.1991.03460110084029 [PubMed] [Google Scholar]
- Armstrong, R. W. , Steinbok, P. , Cochrane, D. D. , Kube, S. D. , Fife, S. E. , & Farrell, K. (1997). Intrathecally administered baclofen for treatment of children with spasticity of cerebral origin. Journal of Neurosurgery, 87, 409–414. https://doi.org/10.3171/jns.1997.87.3.0409 [DOI] [PubMed] [Google Scholar]
- Awaad, Y. , Rizk, T. , Siddiqui, I. , Roosen, N. , McIntosh, K. , & Waines, G. M. (2012). Complications of intrathecal baclofen pump: Prevention and cure. ISRN Neurology, 2012, 575168. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Becker, R. , Alberti, O. , & Bauer, B. L. (1997). Continuous intrathecal baclofen infusion in severe spasticity after traumatic or hypoxic brain injury. Journal of Neurology, 244, 160–166. https://doi.org/10.1007/s004150050067 [DOI] [PubMed] [Google Scholar]
- Becker, R. , Sure, U. , Petermeyer, M. , & Bertalanffy, H. (1999). Continuous intrathecal baclofen infusion alleviates autonomic dysfunction in patients with severe supraspinal spasticity. Journal of Neurology, Neurosurgery and Psychiatry, 66, 114 https://doi.org/10.1136/jnnp.66.1.114 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Borrini, L. , Bensmail, D. , Thiebaut, J. B. , Hugeron, C. , Rech, C. , & Jourdan, C. (2014). Occurrence of adverse events in long‐term intrathecal baclofen infusion: A 1‐year follow‐up study of 158 adults. Archives of Physical Medicine and Rehabilitation, 95, 1032–1038. https://doi.org/10.1016/j.apmr.2013.12.019 [DOI] [PubMed] [Google Scholar]
- Coffey, R. J. , & Burchiel, K. (2002). Inflammatory mass lesions associated with intrathecal drug infusion catheters: Report and observations on 41 patients. Neurosurgery, 50, 78–86. [DOI] [PubMed] [Google Scholar]
- Coffey, R. J. , Cahill, D. , Steers, W. , Park, T. S. , Ordia, J. , Meythaler, J. , … Smith, R. (1993). Intrathecal baclofen for intractable spasticity of spinal origin: Results of a long‐term multicenter study. Journal of Neurosurgery, 78, 226–232. https://doi.org/10.3171/jns.1993.78.2.0226 [DOI] [PubMed] [Google Scholar]
- Dario, A. , Di Stefano, M. G. , Grossi, A. , Casagrande, F. , & Bono, G. (2002). Long‐term intrathecal Baclofen infusion in supraspinal spasticity of adulthood. Acta Neurologica Scandinavica, 105, 83–87. https://doi.org/10.1034/j.1600-0404.2002.1o042.x [DOI] [PubMed] [Google Scholar]
- Guillaume, D. , Van Havenbergh, A. , Vloeberghs, M. , Vidal, J. , & Roeste, G. (2005). A clinical study of intrathecal baclofen using a programmable pump for intractable spasticity. Archives of Physical Medicine and Rehabilitation, 86, 2165–2171. https://doi.org/10.1016/j.apmr.2005.05.018 [DOI] [PubMed] [Google Scholar]
- Kolaski, K. , & Logan, L. R. (2007). A review of the complications of intrathecal baclofen in patients with cerebral palsy. NeuroRehabilitation, 22, 383–395. [PubMed] [Google Scholar]
- Meythaler, J. M. , DeVivo, M. J. , & Hadley, M. (1996). Prospective study on the use of bolus intrathecal baclofen for spastic hypertonia due to acquired brain injury. Archives of Physical Medicine and Rehabilitation, 77, 461–466. https://doi.org/10.1016/S0003-9993(96)90034-9 [DOI] [PubMed] [Google Scholar]
- Meythaler, J. M. , Guin‐Renfroe, S. , Grabb, P. , & Hadley, M. N. (1999). Long‐term continuously infused intrathecal baclofen for spastic‐dystonic hypertonia in traumatic brain injury: 1‐year experience. Archives of Physical Medicine and Rehabilitation, 80, 13–19. https://doi.org/10.1016/S0003-9993(99)90301-5 [DOI] [PubMed] [Google Scholar]
- Meythaler, J. M. , McCary, A. , & Hadley, M. N. (1997). Prospective assessment of continuous intrathecal infusion of baclofen for spasticity caused by acquired brain injury: A preliminary report. Journal of Neurosurgery, 87, 415–419. https://doi.org/10.3171/jns.1997.87.3.0415 [DOI] [PubMed] [Google Scholar]
- Motta, F. , & Antonello, C. E. (2014). Analysis of complications in 430 consecutive pediatric patients treated with intrathecal baclofen therapy: 14‐year experience. Journal of Neurosurgery Pediatrics, 13, 301–306. https://doi.org/10.3171/2013.11.PEDS13253 [DOI] [PubMed] [Google Scholar]
- Motta, F. , & Antonello, C. E. (2016). Comparison between an Ascenda and a silicone catheter in intrathecal baclofen therapy in pediatric patients: Analysis of complications. Journal of Neurosurgery Pediatrics, 18, 493–498. https://doi.org/10.3171/2016.4.PEDS15646 [DOI] [PubMed] [Google Scholar]
- Motta, F. , Buonaguro, V. , & Stignani, C. (2007). The use of intrathecal baclofen pump implants in children and adolescents: Safety and complications in 200 consecutive cases. Journal of Neurosurgery, 107(1 Suppl), 32–35. [DOI] [PubMed] [Google Scholar]
- Paskhin, D. L. , Dekopov, A. V. , Tomsky, A. A. , Isagulyan, E. D. , & Salova, E. M. (2017). Complications of intrathecal baclofen therapy. Zhurnal Voprosy Neirokhirurgii Imeni N. N. Burdenko, 81, 63–69. https://doi.org/10.17116/neiro201780763-69 [DOI] [PubMed] [Google Scholar]
- Penn, R. D. , & Kroin, J. S. (1985). Continuous intrathecal baclofen for severe spasticity. Lancet, 2, 125–127. https://doi.org/10.1016/S0140-6736(85)90228-4 [DOI] [PubMed] [Google Scholar]
- Penn, R. D. , Savoy, S. M. , Corcos, D. , Latash, M. , Gottlieb, G. , Parke, B. , & Kroin, J. S. (1989). Intrathecal baclofen for severe spinal spasticity. New England Journal of Medicine, 320, 1517–1521. https://doi.org/10.1056/NEJM198906083202303 [DOI] [PubMed] [Google Scholar]
- Rawicki, B. (1999). Treatment of cerebral origin spasticity with continuous intrathecal baclofen delivered via an implantable pump: Long‐term follow‐up review of 18 patients. Journal of Neurosurgery, 91, 733–736. https://doi.org/10.3171/jns.1999.91.5.0733 [DOI] [PubMed] [Google Scholar]
- Rawlins, P. K. (2004). Intrathecal baclofen therapy over 10 years. Journal of Neuroscience Nursing, 36, 322–327. https://doi.org/10.1097/01376517-200412000-00005 [DOI] [PubMed] [Google Scholar]
- Rifici, C. , Kofler, M. , Kronenberg, M. , Kofler, A. , Bramanti, P. , & Saltuari, L. (1994). Intrathecal baclofen application in patients with supraspinal spasticity secondary to severe traumatic brain injury. Functional Neurology, 9, 29–34. [PubMed] [Google Scholar]
- Stetkarova, I. , Yablon, S. A. , Kofler, M. , & Stokic, D. S. (2010). Procedure‐ and device‐related complications of intrathecal baclofen administration for management of adult muscle hypertonia: A review. Neurorehabilitation and Neural Repair, 24, 609–619. https://doi.org/10.1177/1545968310363585 [DOI] [PubMed] [Google Scholar]
- Turner, J. A. , Sears, J. M. , & Loeser, J. D. (2007). Programmable intrathecal opioid delivery systems for chronic noncancer pain: A systemic review of effectiveness and complications. Clinical Journal of Pain, 23, 180–195. https://doi.org/10.1097/01.ajp.0000210955.93878.44 [DOI] [PubMed] [Google Scholar]
- Veizi, I. E. , Hayek, S. M. , Hanes, M. , Galica, R. , Katta, S. , & Yaksh, T. (2016). Primary hydromorphone‐related intrathecal Catheter tip granulomas: Is there a role for dose and concentration? Neuromodulation: Technology at the Neural Interface, 19, 760–769. https://doi.org/10.1111/ner.12481 [DOI] [PubMed] [Google Scholar]
- Yaksh, T. L. , Hassenbusch, S. , Burchiel, K. , Hildebrand, K. R. , Page, L. M. , & Coffey, R. J. (2002). Inflammatory masses associated with intrathecal drug infusion: A review of preclinical evidence and human data. Pain Medicine, 3, 300–312. https://doi.org/10.1046/j.1526-4637.2002.02048.x [DOI] [PubMed] [Google Scholar]