Table 1.

Imaging settings and effective doses for the Siemens scanner

No. Arms mAs Scout Strength ED (mSv)
1 Down 50 PA + Lat Average 3.2
2 Down 50 PA Average 4.2
3 Down 50 4.9
4 Down 25 PA + Lat Average 1.6
5 Down 25 PA Average 2.1
6 Down 100 PA + Lat Average 6.6
7 Down 100 PA Average 8.6
8 Down 50 AP + Lat Average 3.2
9 Down 50 AP Average 3.3
10 Down 50 Lat Average 3.4
11 Down 50 PA + Lat Weak 3.4
12 Down 50 PA Weak 4.0
13 Down 50 PA + Lat Strong 3.4
14 Down 50 PA Strong 5.2
15 Up 50 PA + Lat Average 2.0
16 Up 50 PA Average 4.1
17 Up 50 4.9
18 Up 25 PA + Lat Average 1.0
19 Up 25 PA Average 1.9
20 Up 100 PA + Lat Average 4.2
21 Up 100 PA Average 8.4

mAs = quality reference mAs when AEC was applied; Scout = direction of the scout image(s) used for dose modulation; Strength = strength of dose modulation; ED = effective dose. Only mean values are presented for ED because all SDs are 0.0.