Skip to main content
. Author manuscript; available in PMC: 2018 Dec 29.
Published in final edited form as: Phys Med Biol. 2017 Dec 29;63(1):015031. doi: 10.1088/1361-6560/aa9874

Figure 3.

Figure 3

(a) Fluorescence SVD amplitude for Photofrin in tissue-simulating phantom experiments with different optical properties (μa = 0.1 – 0.9 cm−1 and μs = 5 – 24 cm−1) but constant Photofrin concentration of 3mg/kg. Photofrin SVD amplitudes were corrected for optical properties using Eq. (3) to obtained corrected SVD amplitude, Ap,ref (b) Fluorescence spectra of Photofrin at concentrations ranging from 0.0625 mg/kg to 9 mg/kg in tissue simulating phantom with μa,ref = 0.3 cm−1 and μs,ref = 9.6 cm−1 and (c) Photofrin concentration calibration curve.