Table 2: Percentage of matched continuous glucose monitoring reference pairs in specified glucose ranges from US Food and Drug Administration summaries of device safety and effectiveness data.

CGM Glucose Range mg/dL (mmol/L) rtCGM*† isCGM
Dexcom39 Medtronic US40 Abbott41
<40 (<2.2) 6.0% 0.0% 17%
≥40-60 (≥2.2-3.3) 74% 37% 23%
>60-80 (>3.3-4.4) 68% 64% 37%
>80-120 (>4.4-6.7) 72% 55% 70%
>120-160 (>6.7-8.9) 72% 55% 64%
>160-200 (>8.9-11.1) 59% 50% 58%
>200-250 (>11.1-13.9) 70% 53% 54%
>250-300 (>13.9-16.7) 61% 54% 47%

*Dexcom G4 system, performed with the Software 505, which is used in the Dexcom G5 Mobile system. †Calibration was performed every 12 hours. CGM = continous glucose monitoring; isCGM = intermittently scanned continuous glucose monitoring; rtCGM = real-time continuous glucose monitoring.