Table 1.
Second generation anatomic 3D modeling QA phantom. Specific test objects and their dimensions are described and are also illustrated in Fig. 13
| Test object | Test | Dimensions |
|---|---|---|
| Line pair pattern | Spatial resolution | Line pairs(lp) at 0.5, 1.0, and 2.0 lp/mm along each physical axis |
| Concave/Convex surface | Curvilinear surface accuracy | Height-7.5 mm Length-50 mm Width-12.5 mm Depth-10.0 mm |
| Spheres | Shape fidelity | Semi spheres at diameters of 2.5 mm, 5.0 mm, and 7.5 mm |
| Cones | Shape fidelity | Positive cone: Bottom diameter-20 mm Top diameter-10 mm Height-20 mm Negative cone: Top diameter-16.5 mm Bottom diameter-10 mm Height-7.5 mm |
| Hexagonal Cylinder | Rectilinear geometric accuracy | Width-10 mm Height-20 mm |
| Positive base | Volumetric accuracy & temporal stability | Width-50 mm Length-100 mm Height-5 mm |
| Negative base | Volumetric accuracy & temporal stability | Width-50 mm Length-100 mm Height-12.5 mm |
| Ruler | Spatial accuracy | Five markers placed at 10 mm spacing along physical X and Y edges of phantom on negative base. |
| Spiral | Anatomic | Inner diameter-1.8 mm Outer diameter-2.5 mm Spiral start-end separation-10 mm |
| Vascular tree | Anatomic | Cylinders of uniform length of 10 mm with diameters of 5 mm, 2.5 mm, 1.25 mm, 0.75 mm and 0.5 mm. Two outer branches at diameters of 2.5 mm and 0.75 mm placed at 45° and 25° respectively to the inner cylinders. |