Skip to main content
. 2018 May 2;2018:8734013. doi: 10.1155/2018/8734013

Table 5.

Soxhlet extraction-GC/MS method validation parameters for the 16 EPA PAHs in SRM 1649b (n=7).

PAH Experimental mean (ng)a NIST-certified value (ng)a RSD (%) Recovery (%) Accuracyb (%)
Naph 67.3 ± 18.6 391 ± 35 33.8 17.2 −82.8
Acy 20.2 ± 5.9 79.9 ± 9.5 35.9 25.2 −74.8
Ace 25.0 ± 5.4 81.6 ± 16.5 26.5 30.7 −69.4
FL 32.6 ± 9.2 92.3 ± 14.4 34.5 35.3 −64.7
Phe 1215 ± 331 1668 ± 24 33.4 72.8 −27.2
Ant 104 ± 25 169 ± 1 32.5 61.8 −38.2
Ft 2392 ± 559 2573 ± 32 31.0 93.0 −7.05
Pyr 1914 ± 398 2054 ± 57 27.6 93.2 −6.79
BaA 808 ± 148 870 ± 20 24.3 92.9 −7.13
Chry 1632 ± 464 1256 ± 11 37.6 129 29.9
BbFt + BjFt 3144 ± 800 3260 ± 91 33.7 96.4 −3.58
BkFt 921 ± 319 702 ± 20 45.9 131 31.3
BaP 1019 ± 271 1018 ± 98 35.2 100 0.07
IP 1109 ± 296 1192 ± 65 35.4 93.1 −6.89
DBahA + DBacA 507 ± 176 241 ± 4 45.9 363 263
BghiP 2143 ± 580 1777 ± 32 35.8 120 20.6
Averagec 34.9 120 |36.8|

aExpanded uncertainty about the mean, with coverage factor, k = 2; baccuracy = (experimental value − certified value) × 100/certified value; cexcept Naph, Acy, Ace, and FL.