Skip to main content
. 2018 May 16;13(5):e0197074. doi: 10.1371/journal.pone.0197074

Table 3. Variation between replicate, extraction, assay run, and laboratory observed during repeatability and reproducibility studies.

Replicate Extraction Assay Run Laboratory
LN34 Assay CV 0.35 ± 0.06% 0.72 ± 0.27% 1.33 ± 0.32% 9.13 ± 2.06%
SD 0.09 ± 0.02 0.12 ± 0.07 0.36 ± 0.16 1.89 ± 0.45
ΔCt 0.18 ± 0.04 0.39 ± 0.12 0.99 ± 0.43 3.70 ± 0.94
β-Actin Assay CV 0.48 ± 0.13% 2.15 ± 0.75% 1.86 ± 0.70% 6.56 ± 1.00%
SD 0.10 ± 0.03 0.44 ± 0.12 0.41 ± 0.10 1.34 ± 0.20
ΔCt 0.18 ± 0.05 0.97 ± 0.28 1.14 ± 0.41 2.57 ± 0.39

Variability shown as average coefficient of variation (CV), standard deviation (SD) and difference in Ct values (ΔCt) with 95% confidence intervals.