Table 3. Variation between replicate, extraction, assay run, and laboratory observed during repeatability and reproducibility studies.
Replicate | Extraction | Assay Run | Laboratory | ||
---|---|---|---|---|---|
LN34 Assay | CV | 0.35 ± 0.06% | 0.72 ± 0.27% | 1.33 ± 0.32% | 9.13 ± 2.06% |
SD | 0.09 ± 0.02 | 0.12 ± 0.07 | 0.36 ± 0.16 | 1.89 ± 0.45 | |
ΔCt | 0.18 ± 0.04 | 0.39 ± 0.12 | 0.99 ± 0.43 | 3.70 ± 0.94 | |
β-Actin Assay | CV | 0.48 ± 0.13% | 2.15 ± 0.75% | 1.86 ± 0.70% | 6.56 ± 1.00% |
SD | 0.10 ± 0.03 | 0.44 ± 0.12 | 0.41 ± 0.10 | 1.34 ± 0.20 | |
ΔCt | 0.18 ± 0.05 | 0.97 ± 0.28 | 1.14 ± 0.41 | 2.57 ± 0.39 |
Variability shown as average coefficient of variation (CV), standard deviation (SD) and difference in Ct values (ΔCt) with 95% confidence intervals.