Skip to main content
. 2018 May 16;13(5):e0196683. doi: 10.1371/journal.pone.0196683

Fig 6. Resistance for each system with increased submersion depths and a fresh gas flow of 10 L/min.

Fig 6

There is an increase in pressure during expiration and a decrease during inspiration. The revised Pumani has higher resistance (steeper slope) than conventional bubble CPAP systems. Simulations without submersion were included to illustrate resistance from the connectors, tubing and prongs. Negative flows correspond to patient inspiration.