Fig 7. Resistance for each system with increasing fresh gas flows and a submersion depth of 5.0 cm.

Fig 7

Due to the system resistance, with increasing fresh gas flows, CPAP (pressure at 0 L/min of simulated airway flow) increased slightly in the Diamedica and the Fisher & Paykel and more pronounced in the revised Pumani. For all systems the resistance (slope) was similar at different levels of fresh gas flows. An uncut endotracheal tube size 3 (no fresh gas flow) was included for comparison. Negative flows correspond to patient inspiration.