Skip to main content
. 2018 Jan 22;183(3):513–519. doi: 10.1007/s11046-017-0238-1

Table 3.

Test characteristics of the four antigen assays

Platelia Platelia plus Serion Fungitell
Proven IC Proven, probable or possible IC Proven IC Proven, probable or possible IC Proven IC Proven, probable or possible IC Proven IC Proven, probable or possible IC
True positives 13 29 11 20 12 26 12 34
True negatives 709 709 259 259 710 710 170 170
False negatives (false-negative rate) 7 (35%) 41 (58.5%) 2 (15.4%) 31 (60.8%) 11 (47.8%) 44 (62.9%) 0 (0%) 17 (33.3%)
False positives (false-positive rate) 14 (1.9%) 14 (1.9%) 31 (10.7%) 31 (10.7%) 17 (2.3%) 17 (2.3%) 123 (42%) 123 (42%)
Total 743 793 303 341 750 797 305 344
Sensitivity 0.65 0.41 0.85 0.39 0.52 0.37 1 0.67
Specificity 0.98 0.98 0.89 0.89 0.98 0.98 0.58 0.58
PPV 0.48 0.67 0.26 0.39 0.41 0.60 0.09 0.22
NPV 0.99 0.95 0.99 0.89 0.98 0.94 1 0.91