. 2018 May 15;9:911. doi: 10.3389/fmicb.2018.00911

Table 1.

Global distribution of identification results with rapid automated method using Blood Culture Sample Preparation (BCSP) prototype and identification by Vitek®MS system.

Identification rate (%)
Type of inoculated microorganism Reference: 18–48 h subculture + Vitek®MS identification Direct identification from positive blood culture using BCSP prototype with BacT/ALERT 3D, SA bottles, and Vitek®MS identification Direct identification from positive blood culture using BCSP prototype with BacT/ALERT Virtuo, FA Plus bottles, and Vitek®MS identification
Gram-negative species 96 % [91–99]% 85% [79–89]% 86% [71–95]%
Gram-positive species 98 % [93–100]% 88% [82–92]% 86% [71–95]%
Yeast species 100 % [81–100%] 100% [90–100]% 75% [48–93]%
Overall ID rate 97.3% [94–99]% 87% [84–90]% 84% [75–91]%

SA and FA Plus bottles were inoculated with microorganisms belonging to strain collection of the study and were incubated in BacT/ALERT 3D® and BacT/ALERT Virtuo® systems, respectively. Results were compared with reference method based subculture of positive blood culture bottle (COS plate) and identification by Vitek®MS system. The results in square brackets are the 95% two-sided exact confidence interval.