Table 1.
Troubleshooting table.
Problem | Possible reason | Solution |
---|---|---|
Embryos are carried away during inserting. | Embryos are glued not enough. | Prepare thicker heptane glue to stable embryos. |
The tip is not sharp enough. | Adjust the tip size of micropipettes during preparation. | |
Body fluids flow out during injection. | Not enough Dehydration. | Prolong dehydration to 6-7 min. |
The flow from the micropipette tip is too high. | Use the micropipette with an appropriate tip size. | |
Reduce the output pressure or the pulse duration. | ||
Dextran-positive macropinosomes cannot be seen after 30 min of injection. | The micropipette tip is sealed or its tip is too small. | Replace the micropipette with a newly-made one. |
Use the micropipette with an appropriate tip size. | ||
Air bubbles are trapped in the micropipette tip. | Gently flick the micropipette to discharge bubbles or reload the injection solution. | |
Large dextrans are filtered during diffusion. | Use small dextrans. 3 kDa-dextran is recommended. | |
Embryos are not healthy enough. | Use another healthy embryo. | |
Operate embryos as quickly as possible. | ||
High background in the extracellular space | The micropipette tip is too large. | Adjust the tip size of micropipettes. |
The output pressure is too high or pulse duration is too long. | Reduce the output pressure or the pulse duration. |