Embryos are carried away during inserting. |
Embryos are glued not enough. |
Prepare thicker heptane glue to stable embryos. |
|
The tip is not sharp enough. |
Adjust the tip size of micropipettes during preparation. |
Body fluids flow out during injection. |
Not enough Dehydration. |
Prolong dehydration to 6-7 min. |
|
The flow from the micropipette tip is too high. |
Use the micropipette with an appropriate tip size. |
|
|
Reduce the output pressure or the pulse duration. |
Dextran-positive macropinosomes cannot be seen after 30 min of injection. |
The micropipette tip is sealed or its tip is too small. |
Replace the micropipette with a newly-made one. |
|
|
Use the micropipette with an appropriate tip size. |
|
Air bubbles are trapped in the micropipette tip. |
Gently flick the micropipette to discharge bubbles or reload the injection solution. |
|
Large dextrans are filtered during diffusion. |
Use small dextrans. 3 kDa-dextran is recommended. |
|
Embryos are not healthy enough. |
Use another healthy embryo. |
|
|
Operate embryos as quickly as possible. |
High background in the extracellular space |
The micropipette tip is too large. |
Adjust the tip size of micropipettes. |
|
The output pressure is too high or pulse duration is too long. |
Reduce the output pressure or the pulse duration. |