Table 3.

Scan parameters and CTDI of 3 initial CBCT protocols

  Head Pelvis Thorax
Tube voltage (kVp) 100 125 110
Tube current (mA) 20 42 20
Exposure time per projection (ms) 20 20 20
C-ring rotation range (°) 200 360 360
Frame rate (fps) 10 10 10
Number of projections 340 600 600
Fan mode Full Half Half
Reconstruction matrix 512 × 512 1024 × 1024 1024 × 1024
Bow-tie filter Full Half Half
Pixel size (mm) 0.55 0.52 0.52
Slice spacing (mm) 0.6 1 1
Phantom diameter (cm) 16 32 32
CTDI (centre) (mGy) 3.4 6.5 2.0
CTDI (top) (mGy) 1.9 9.7 3.2
CTDI (right) (mGy) 4.2 9.3 3.1
CTDI (bottom) (mGy) 5.1 8.8 2.9
CTDI (left) (mGy) 3.5 11.5 3.8
Weighted CTDI (mGy) 3.6 8.7 2.9

CBCT, cone beam CT; CTDI, CT dose index.