Table 1.
Clinical coronary CTA scan protocol on 2nd and 3rd generation DSCT
| 2nd generation DSCT | 3rd generation DSCT | |
|---|---|---|
| Topogram – Lateral and PA | ||
| Tube voltage/tube current | 120 kVp/30 mA | 120 kVp/30 mA |
| Noncontrast scan for calcium scoring and range planning | ||
| Tube voltage | 120 kV (fixed) | 120 kV (fixed) |
| Tube current | reference mA 80 (CareDose 4D enabled) | reference mA 80 (CareDose 4D enabled) |
| Triggering/gating | Prospectively ECG-triggered axial-sequential, single phase at absolute delay of 280 msec | Prospectively ECG-triggered axial-sequential, single phase at absolute delay of 280 msec |
| Kernel | Filtered back projection B35f | Filtered back projection Qr36 |
| Rotation time | 0.28 sec | 0.25 sec |
| Coronary CTA | ||
| Tube voltage | reference kV 120 (Care kV enabled) | reference kV 120 (Care kV enabled) |
| Tube current | reference mA 280 (CareDose 4D enabled) | reference mA 280 (CareDose 4D enabled) |
| ECG synchronization | ||
| Prospectively ECG triggered (heart rate < approximately 75 beats per min or arrhythmia) |
|
|
| Retrospectively ECG gated (heart rate > approximately 75 beats per min and regular; or difficulty breath holding) |
|
|
| Tube current modulation |
|
|
| Contrast timing | Isovue 370, bolus timing method with 20cc test bolus | Isovue 370, bolus timing method with 20cc test bolus |
| Contrast injection | Power injector with mixed contrast injection weight-based software | Power injector with mixed contrast injection weight-based software |
| Kernel/Data Reconstruction | I131F SAFIRE strength 3 (out of 5) | Bv40 ADMIRE strength 3 (out of 5) |
| Rotation time | 0.28 sec | 0.25 sec |
| Slice thickness | 0.6 mm | 0.6 mm |
| Slice increment | 0.3 mm | 0.3 mm |